50 TX6356-UM-EN-
01
6.2.3 Sentro Humidity Sensor
- Clean
1. Examine the sensor housing and assess
its condition.
2. Clean the sensor head with a soft brush
or cloth if necessary.
Checkpoint
Do not use sharp tools as this may cause
damage to the humidity sensor.
3. After the completion of all maintenance,
update the maintenance records.
6.2.4 Sentro Humidity Sensor
Output Signal - Check
1. Using a test meter check that the value
of the output signal agrees with the
value of the display reading.
2. After the completion of all maintenance,
update the maintenance records.