Trio CS-1566A Instruction Manual Download Page 20

S I N G L E - C H A N N E L  A P P L I C A T I O N S 

introduction. 

In addition to the dual-trace applications previously 

outlined, there are, of course, many servicing and 

laboratory applications where only single-trace or 

single-channel operation of the oscilloscope is re-

quired. By setting the MODE switch to CH2 and 

using the  C H 2 amplifier, many flexible operations 

will be achieved; and, in addition, by placing the 

M O D E switch to  A D D and the  C H 2 polarity switch 

to the  P U L L  I N V E R T position. Whatever waveform 

is obtained can be inverted in polarity if desired by 

the operator. 

T e l e v i s i o n  S e r v i c i n g : 

A triggered sweep oscilloscope is advantageous in 

servicing and aligning television receivers. This 

oscilloscope also includes several features that 

were incorporated to make television servicing and 

V I D E O signal observation easier and more 

comprehensive. 

These feature inc ude: 

* With the  S Y N C switch set to  V I D E O position, 

the  S W E E P  T I M E / D I V control automatically 

selects the  V I D E O - F R A M E sync at sweep 

speeds appropriate for viewing frame and 

V I D E O - L I N E sync at sweep speeds appropriate 

for viewing lines. 

* Wide bandwidth for high resolution video and 

high speed pulse presentation. 

S i n g l e - t r a c e Operation and  P e a k - t o - P e a k 

Voltage  R e a d i n g s : 

For general troubleshooting and isolation of 

troubles in almost any electronic equipment, the 

oscilloscope is an indispensable instrument. It 

provides a visual display of the absence or presence 

of normal signals. This method (signal-tracing) 

may be used to trace a signal by measuring several 

points in the signal path.  A s measurements 

proceed along the signal path, a point may be 

found where the signal disappears. When this 

happens, the source of trouble has been located. 

However, the oscilloscope shown much more than 

the mere presence or absence of signals. It 

provides a peak-to-peak voltage measurement. 

The schematic diagram or accompanying service 

data on the equipment being serviced usually 

includes waveform pictures. There waveform 

picture include the required sweep time and the 

normal peak-to-peak voltage. Compare the 

peak-to-peak voltage readings on the oscilloscope 

with those shown on the waveform pictures. Any 

abnormal readings should be followed by additional 

readings in the suspected circuits until the trouble 

is isolated to as small an area as possible. 

Composite Video  W a v e f o r m  A n a l y s i s 

Probably the most important  w a v e f o r m in 

television servicing is the composite waveform 

consisting of the video signal, the blanking pedestal 

signal and the sync pulses.

 Fig. 19

 and

 Fig.  2 0 

show typical oscilloscope traces when observing 

composite video signals synchronized with ho-

rizontal sync pulses and vertical blanking pulses. 

Composite video signals can be observed at various 

stages of the television receiver to determine 

w h e t h e r  c i r c u i t s are  p e r f o r m i n g  n o r m a l l y . 

Knowledge of waveform makeup, the appearance 

of a normal waveform, and the cause of various ab-

normal waveforms help the technician locate and 

correct many problems. The technician should 

study such waveforms in a television receiver 

known to be in good operating condition, noting 

the waveform at various points in the video 

amplifier. To set up the oscilloscope for viewing 

composite video waveforms, use the following 

procedures: 

1. Turn the television set to a local channel. A 

test tape or a signal generator also can be used 

for service work. 

2. Set the  M O D E switch to  C H 2 position. 

3. Set the  S W E E P  T I M E / D I V switch to the 10

 us 

position for observing  T V horizontal lines or to 

the 2 ms position for observing TV vertical 

frames. 

4. Set the  S Y N C switch to  V I D E O +  o r V I D E O -

position. 

5. Set the  S O U R C E switch to INT position. 

6. Best overall sync performance is obtained 

when the  T R I G  L E V E L / P U L L  A U T O control is 

pushed in. It may be pulled out initially to 

provide continuous sweep during set-up. 

7. Set the  C H 2  A C - G N D - D C switch to the  A C 

position. 

8. Connect the ground clip of the probe to the 

c h a s s i s . With the probe set for 10 : 1 

attenuation, connect the tip of the probe to the 

video detector output. 

9. Set the  C H 2  V O L T S / D I V switch for the largest 

vertical deflection possible without going 

offscale. 

10. If necessary, rotate the  T R I G  L E V E L control to 

a position that provides a synchronized display. 

1 1 . Adjust the sweep time  V A R I A B L E for two ho-

rizontal lines or  t w o vertical  f r a m e s of 

20 

Summary of Contents for CS-1566A

Page 1: ...TRIGGERED SWEEP OSCILLOSCOPE HIGH S T A B I L I T Y CS 1566A DUAL TRACE OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Page 2: ...observation The time base switch allows changeover between V vertical and H horizontal of VIDEO sync separator circuit automatically and electronically INT C H 1 CH2 LINE and EXT can be individually...

Page 3: ...time 0 5Ms div to 5s div 5 and X Y 1 2 5 sequence Fine adjustment in all 19 ranges Magnification 10 times 5 PULL x 10 MAG Linearity Better than 3 2 zs div to 0 5s div Better than 5 0 5 s div to 1 s d...

Page 4: ...tation Trace angle is adjustable by panel surface ad justor Power Requirements Power supply voltage 100 120 220 240V 10 50 60 Hz Power consumption 47W Dimensions Width 260 mm 277 mm Height 190 mm 204...

Page 5: ...CONTROLS ON PANELS FRONT PANEL Fig 1 REAR PANEL Fig 2 5...

Page 6: ...Dual trace operation in which sweep is chopped at approximate 200 kHz rate and switched between Channel 1 and Channel 2 traces Recommended for sweep times of 0 5 s div to 1 ms div ADD The waveforms f...

Page 7: ...tive going sync pulse 19 LEVEL PULL AUTO Sync level adjustment determines points on waveform slope where sweep starts equals most negative point of triggering and equals most positive point of trigger...

Page 8: ...40V be sure to use a 0 5A fuse 29 POWER CONNECTOR For connection of the supplied AC power cord 30 Z AXIS INPUT Intensity modulation terminal T T L logic compatible high logic increase brightness low l...

Page 9: ...djust VOLTS DIV 4 7 and VARIABLE 5 8 so that the pulse amplitude is set to the 0 and 100 scales 2 Turn SWEEP TIME DIV 21 to magnify the rising portion of the waveform as large as possible VARIABLE 22...

Page 10: ...ative going waveform In this case the output waveform is shifted with respect to the leading edge of the Fig 5 Waveforms in divide by two circuit reference frequency pulse by a time interval cor respo...

Page 11: ...of the waveforms change depending upon the input or more mode of opera tion Ffg 8 shows a typical digital circuit and identifies several of the points at which waveform measure ment are appropriate T...

Page 12: ...uch cases it is recommended that the sync re main unchanged while the sweep speed or X 1 0 Fig 8 Typical digital circuit using several time related waveforms MAG used be to expaned the waveform displa...

Page 13: ...e amplifier unit may be measured for distortion with this oscilloscope This type of measurement is especially variable when the slope of a waveform must be faithfully reproduced by an amplifier Fig 10...

Page 14: ...ns Fig 11 shows a burst circuit The basic settings are identical to those of in Fig 5 Waveform EA is the reference waveform and is applied to CH1 input All other waveforms are sampled at CH2 and compa...

Page 15: ...Y P E T O H O R I Z O N T A L A M P L I F I E R T O V I D E O O U T P U T A M P L I F I E R N O T E NO E L E C T R I C A L C O N N E C T I O N P L A C E C L I P ON I N S U L A T I O N O F P L A T E C...

Page 16: ...ut jack The VARIABLE control is adjusted as required to provide a complete cycle of the input waveform displayed on 8 div horizontally A waveform height of 2 div is used The 8 div display represents 3...

Page 17: ...abled to see the VITS The transmitted VITS is precision sequence of a specific frequences amplitudes and waveshapes as shown in Fig 16 Television networks use the precision signals for adjustment and...

Page 18: ...s on either side This suggests an IF trap is detuned into the passband chopping out frequencies about 2 MHz below the picture carrier frequency Switch to another channel carrying VITS If the same thin...

Page 19: ...ection moving the trace to the left until the expanded VITS information appears as shown in Fig 16 Because of the low repetition rate and the high sweep speed combination the brightness level of the s...

Page 20: ...he peak to peak voltage readings on the oscilloscope with those shown on the waveform pictures Any abnormal readings should be followed by additional readings in the suspected circuits until the troub...

Page 21: ...controls for the desired brightness and best focus 15 To view a specific portion of the waveform such as the color burst pull outward on the POSITION control for X10 magnification Rotate the same con...

Page 22: ...tal gain controls for a display similar to that shown in Fig 23A 5 Set the marker generator precisely to 10 7 MHz The marker pip should be in the center of the bandpass 6 Align the IF amplifiers accor...

Page 23: ...e waveform A sine wave input is applied to the audio circuit being tested The same sine wave input is applied to the vertical input of the oscilloscope and the output of the tested circuit is applied...

Page 24: ...ces a circular oscilloscope pattern Phase shift of less or more than 90 produces an elliptical oscilloscope pattern The amount of phase shift can be calculated from the oscilloscope trace as shown in...

Page 25: ...can evaluate the input and output quality of a signal of many frequencies the harmoniques of the square wave which is what the amplifier sees when amplifying complex wave forms of musical instruments...

Page 26: ...ase a deliberate non linear circuit such as a clipper network 3 The third is delay or phase distortion which is distortion produced by a shift in phase between one or more components of a complex wave...

Page 27: ...lts are opposite in the two cases because of the difference in polarity of the phase angle in the two cases that can be checked through algebraic addition of components Fig 36 indicates low frequency...

Page 28: ...the square wave indicates a transient oscillation in a relatively high Q network in the amplifier circuit In this case the sudden transition in the square wave potential from a sharply rising relativ...

Page 29: ...tuated fundamental C High frequency loss No phase shift D Low frequency phase shift E Low frequency loss and phase shift F High frequency loss and low frequency phase shift G High frequency loss and p...

Page 30: ...or EXT TRIG is up to 50V DC AC peak and the input to Z AXIS is up to 50V DC AC peak 4 Do not increase the brightness of the CRT unnecessarily 5 Do not leave the oscilloscope for a long period with a b...

Page 31: ...adjustments use a well insulated screw driver 3 Before marking adjustments be sure to turn on the power and wait until the unit is stabilized 4 For adjustment follow the procedures described below 5...

Page 32: ...the trace comes to the position of the line in CH2 NOR mode when the CH2 polarity is inverted Horizontal Position Adjustment 1 To adjust the horizontal position under the nor mal sweep condition set t...

Page 33: ...SCHEMATIC DIAGRAM CS 1566A 33...

Page 34: ...A p r o d u c t of T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 2 5 3 1 2 PRINTED IN J A P A N B50 2918 00 G...

Reviews: