Trio CS-1566A Instruction Manual Download Page 16

corresponding to the undesired 50 Hz component. 

The  C H 2 display indicates a waveform of equal 

amplitude and identical phase to the average of the 

CH1 waveform. With the  M O D E switch set to 

A D D and the CH1 signal inverted, and by adjusting 

the  C H 2 vertical attenuator control, and  5 0 Hz 

component of the  C H 1 signal can be cancelled by 

the  C H 2 input and the desired waveform can be 

observed. 

Stereo Amplifier Servicing: 

Another convenient use for a dual-trace oscil-

loscope is in troubleshooting stereo amplifiers. If 

identical amplifiers are used and the output of one 

is weak, distorted or otherwise abnormal, the 

dual-trace oscilloscope can be efficiently used to 

localize the defective state. With an identical sig-

nal applied to the inputs of both amplifiers, a 

side-by-side comparison of both units can be made 

by progressively sampling identical signal points in 

both amplifiers. When the defective or malfunctio-

ning stage has been located, the effect of whatever 

troubleshooting and repair methods are employed 

can be observed can be observed and analyzed im-

mediately. 

Amplifier Phase Shift Measurements: 

Phase measurements can be made by several 

method using oscilloscopes. Typical applications 

are in circuits designed to produce a specific phase 

shift, and measurement of phase shift distortion in 

audio amplifiers and network. In all amplifiers, a 

phase shift is always associated with a change in 

amplitude response. For example, at the

 —

 3 dB 

response points, a phase shift of  4 5 ° occurs. 

Phase  m e a s u r e m e n t s can be performed by 

operating the oscilloscope either in the dual-trace 

mode to measure amplifier phase shift directly. 

Fig. 15

 illustrates this method. In this case, the 

measurements are being made at approximately 

5 0 0 0 Hz. The input signal to the audio amplifier is 

used as a reference and is applied to the  C H 1 input 

jack. 

The  V A R I A B L E control is adjusted as required to 

provide a complete cycle of the input waveform 

displayed on 8 div horizontally. A waveform height 

of 2 div is used. The 8 div display represents  3 6 0 ° 

at the displayed frequency and each division 

represents  4 5 ° of the waveform. 

The signal developed accross the output of the 

audio amplifier is applied to the  C H 2 input jack. 

The vertical attenuator controls of  C H 2 are 

adjusted as required to produce a peak-to-peak 

waveform of 2 div amplitude as shown in

 Fig. 

15B.

 The  C H 2 POSITION control is then adjusted 

so that the  C H 2 waveform is displayed on the same 

horizontal axis as the  C H 1 waveform as shown in 

Fig. 15B.

 The distance between corresponding 

points on the horizontal axis for the two waveforms 

then represents the phase shift between the two 

waveforms. In this case, the zero crossover points 

of the two waveforms are compared. It is shown 

that a difference of 1 div exists. This is then inter-

preted as means a phase shift of  4 5 ° . 

Television Servicing: 

Many of the television servicing procedures can be 

performed using single-trace operation. These are 

outlined later in the applications section covering 

single-trace operations. One of these procedures, 

viewing the  V I T S (vertical interval test signal), can 

be accomplished much more effectively using a 

dual-trace oscilloscope.  A s outlined in the 

single-trace applications section and as shown in 

Fig. 16.

 the information on the Field 1 and Field 2 

vertical blanking interval pulse is different. This is 

shown in detail in

 Fig. 16.

 Also, because the oscil-

loscope sweep is synchronized to the vertical 

blanking interval waveform, the Field 1 and Field 2 

waveforms are superimposed onto each other. 

With dual-trace operation, the signal information 

on each blanking pulse can be viewed separately 

without overlapping.

 Fig. 18

 indicates the oscil-

16 

A D J U S T  A S  R E Q U I R E D 

FOR  C O M P L E T E  C Y C L E 

A F 

S I G N A L 

G E N E R A T O R 

A U D I O 

A M P L I F I E R 

O U T P U T 

L O A D 

I N P U T 

Fig. 15 Measurering amplifier phase shift 

Summary of Contents for CS-1566A

Page 1: ...TRIGGERED SWEEP OSCILLOSCOPE HIGH S T A B I L I T Y CS 1566A DUAL TRACE OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Page 2: ...observation The time base switch allows changeover between V vertical and H horizontal of VIDEO sync separator circuit automatically and electronically INT C H 1 CH2 LINE and EXT can be individually...

Page 3: ...time 0 5Ms div to 5s div 5 and X Y 1 2 5 sequence Fine adjustment in all 19 ranges Magnification 10 times 5 PULL x 10 MAG Linearity Better than 3 2 zs div to 0 5s div Better than 5 0 5 s div to 1 s d...

Page 4: ...tation Trace angle is adjustable by panel surface ad justor Power Requirements Power supply voltage 100 120 220 240V 10 50 60 Hz Power consumption 47W Dimensions Width 260 mm 277 mm Height 190 mm 204...

Page 5: ...CONTROLS ON PANELS FRONT PANEL Fig 1 REAR PANEL Fig 2 5...

Page 6: ...Dual trace operation in which sweep is chopped at approximate 200 kHz rate and switched between Channel 1 and Channel 2 traces Recommended for sweep times of 0 5 s div to 1 ms div ADD The waveforms f...

Page 7: ...tive going sync pulse 19 LEVEL PULL AUTO Sync level adjustment determines points on waveform slope where sweep starts equals most negative point of triggering and equals most positive point of trigger...

Page 8: ...40V be sure to use a 0 5A fuse 29 POWER CONNECTOR For connection of the supplied AC power cord 30 Z AXIS INPUT Intensity modulation terminal T T L logic compatible high logic increase brightness low l...

Page 9: ...djust VOLTS DIV 4 7 and VARIABLE 5 8 so that the pulse amplitude is set to the 0 and 100 scales 2 Turn SWEEP TIME DIV 21 to magnify the rising portion of the waveform as large as possible VARIABLE 22...

Page 10: ...ative going waveform In this case the output waveform is shifted with respect to the leading edge of the Fig 5 Waveforms in divide by two circuit reference frequency pulse by a time interval cor respo...

Page 11: ...of the waveforms change depending upon the input or more mode of opera tion Ffg 8 shows a typical digital circuit and identifies several of the points at which waveform measure ment are appropriate T...

Page 12: ...uch cases it is recommended that the sync re main unchanged while the sweep speed or X 1 0 Fig 8 Typical digital circuit using several time related waveforms MAG used be to expaned the waveform displa...

Page 13: ...e amplifier unit may be measured for distortion with this oscilloscope This type of measurement is especially variable when the slope of a waveform must be faithfully reproduced by an amplifier Fig 10...

Page 14: ...ns Fig 11 shows a burst circuit The basic settings are identical to those of in Fig 5 Waveform EA is the reference waveform and is applied to CH1 input All other waveforms are sampled at CH2 and compa...

Page 15: ...Y P E T O H O R I Z O N T A L A M P L I F I E R T O V I D E O O U T P U T A M P L I F I E R N O T E NO E L E C T R I C A L C O N N E C T I O N P L A C E C L I P ON I N S U L A T I O N O F P L A T E C...

Page 16: ...ut jack The VARIABLE control is adjusted as required to provide a complete cycle of the input waveform displayed on 8 div horizontally A waveform height of 2 div is used The 8 div display represents 3...

Page 17: ...abled to see the VITS The transmitted VITS is precision sequence of a specific frequences amplitudes and waveshapes as shown in Fig 16 Television networks use the precision signals for adjustment and...

Page 18: ...s on either side This suggests an IF trap is detuned into the passband chopping out frequencies about 2 MHz below the picture carrier frequency Switch to another channel carrying VITS If the same thin...

Page 19: ...ection moving the trace to the left until the expanded VITS information appears as shown in Fig 16 Because of the low repetition rate and the high sweep speed combination the brightness level of the s...

Page 20: ...he peak to peak voltage readings on the oscilloscope with those shown on the waveform pictures Any abnormal readings should be followed by additional readings in the suspected circuits until the troub...

Page 21: ...controls for the desired brightness and best focus 15 To view a specific portion of the waveform such as the color burst pull outward on the POSITION control for X10 magnification Rotate the same con...

Page 22: ...tal gain controls for a display similar to that shown in Fig 23A 5 Set the marker generator precisely to 10 7 MHz The marker pip should be in the center of the bandpass 6 Align the IF amplifiers accor...

Page 23: ...e waveform A sine wave input is applied to the audio circuit being tested The same sine wave input is applied to the vertical input of the oscilloscope and the output of the tested circuit is applied...

Page 24: ...ces a circular oscilloscope pattern Phase shift of less or more than 90 produces an elliptical oscilloscope pattern The amount of phase shift can be calculated from the oscilloscope trace as shown in...

Page 25: ...can evaluate the input and output quality of a signal of many frequencies the harmoniques of the square wave which is what the amplifier sees when amplifying complex wave forms of musical instruments...

Page 26: ...ase a deliberate non linear circuit such as a clipper network 3 The third is delay or phase distortion which is distortion produced by a shift in phase between one or more components of a complex wave...

Page 27: ...lts are opposite in the two cases because of the difference in polarity of the phase angle in the two cases that can be checked through algebraic addition of components Fig 36 indicates low frequency...

Page 28: ...the square wave indicates a transient oscillation in a relatively high Q network in the amplifier circuit In this case the sudden transition in the square wave potential from a sharply rising relativ...

Page 29: ...tuated fundamental C High frequency loss No phase shift D Low frequency phase shift E Low frequency loss and phase shift F High frequency loss and low frequency phase shift G High frequency loss and p...

Page 30: ...or EXT TRIG is up to 50V DC AC peak and the input to Z AXIS is up to 50V DC AC peak 4 Do not increase the brightness of the CRT unnecessarily 5 Do not leave the oscilloscope for a long period with a b...

Page 31: ...adjustments use a well insulated screw driver 3 Before marking adjustments be sure to turn on the power and wait until the unit is stabilized 4 For adjustment follow the procedures described below 5...

Page 32: ...the trace comes to the position of the line in CH2 NOR mode when the CH2 polarity is inverted Horizontal Position Adjustment 1 To adjust the horizontal position under the nor mal sweep condition set t...

Page 33: ...SCHEMATIC DIAGRAM CS 1566A 33...

Page 34: ...A p r o d u c t of T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 2 5 3 1 2 PRINTED IN J A P A N B50 2918 00 G...

Reviews: