Trio CS-1566A Instruction Manual Download Page 11

Propagation  D e l a y  T i m e  M e a s u r e m e n t : 

A n example of propagation delay in a divide-by-

eight circuit  w a s given in the previous paragraph. 

Significant propagation delay may occur in any cir-

cuit with several consecutive stages. This os-

cilloscope has features which simplify measure-

ment of propagation delay time.  F i g . 7 shows the 

resultant waveforms when the dual-trace presenta-

tion is combined into a single-trace presentation by 

selecting the  A D D position of the MODE switch. 

With  C H 2 PULL. INVERT switch in the normal posi-

tion (pushed in) the two inputs are algebraically 

added in a single-trace display. Similarly, in the in-

verted position (pulled out) the two inputs are 

algebraically subtracted. Either position provides a 

precise display of the propagation time (Tp). Using 

procedures given for calibrated time measurement. 

Fig. 7 Using  A D D mode propagation 

time measurement 

Tp can be measured. A more precise measurement 

can be obtained if the Tp portion of the waveform is 

expanded horizontally. This may be done by pulling 

the  X 1 0  M A G control. It also may be possible to 

view the desired portion of the waveform at a faster 

sweep speed. 

Digital Circuit Time Delay Measurement: 

A dual-trace oscilloscope is a necessity in design-

ing, manufacturing and servicing digital equip-

ment. A dual-trace oscilloscope permits easy com-

parison of time relationships between  t w o 

waveforms. In digital equipment, it is common for 

a large number of circuits to be synchronized, or to 

have a specific time relationship to each other. 

Many of the circuits are frequency dividers as 

priviously described, but waveforms are of ten 

time-related in many other combinations. In the 

dynamic state, some of the waveforms change, 

depending upon the input or more mode of opera-

tion. 

Ffg. 8 shows a typical digital circuit and identifies 

several of the points at which waveform measure-

ment are appropriate. The accompanying

 Fig.

 8 

shows the normal waveforms to be expected at 

each of these points and their timing relationships. 

The individual waveforms have limited value unless 

their timing relationship to one or more of the other 

waveforms is known to be correct. The dual-trace 

oscilloscope allows this comparison to be made. In 

typical fashion, waveform No. 3 would be displayed 

on  C H 1 and waveform No. 4 through No. 8 and No. 

10, would be displayed on  C H 2 although other tim-

ing comparisons may be desired. Waveforms No. 

11 through No. 13 would probably be displayed on 

C H 2 . 

In the family of time-related waveforms shown in 

Fig. 9,

 waveform No. 8 or No. 10 is excellent sync 

source for viewing all of the waveforms; there is 

but one triggering pulse per frame. For con-

venience, external sync, any of the waveforms may 

be displayed without readjustment of the sync con-

trols. 

With No. 8 or No. 10 used as external triggering 

source, any of the waveforms may be displayed 

without readjustment of the  T R I G  L E V E L control. 

Waveforms No. 4 through No. 7 should not be used 

as the triggering source because they do not con-

tain a triggering pulse at the start of the frame. It 

would not be necessary to view the entire 

waveforms as shown in

 Fig. 9

 in all cases. In fact, 

there are many times when a closer examination of 

a portion of the waveforms would be appropriate. 

E X P A N D  T H I S  P O R T I O N 

F O R  T I M E  M O R E  P R E C I S E 

M E A S U R E M E N T 

11 

A REFERENCE FREQUENCY PULSE TRAIN 

(1000 PULSES PER SECOND) 

B

 IDEAL DIVIDE BY EIGHT OUTPUT 

C PROPAGATION DELAY IN DIVIDE BY EIGHT CIRCUIT 

Fig. 6 Waveforms in divide-by-eight circuit 

CH2  P O L A R I T Y 

N O R M A L 

CH2  P O L A R I T Y 

I N V E R T 

Summary of Contents for CS-1566A

Page 1: ...TRIGGERED SWEEP OSCILLOSCOPE HIGH S T A B I L I T Y CS 1566A DUAL TRACE OSCILLOSCOPE INSTRUCTION MANUAL T R I O...

Page 2: ...observation The time base switch allows changeover between V vertical and H horizontal of VIDEO sync separator circuit automatically and electronically INT C H 1 CH2 LINE and EXT can be individually...

Page 3: ...time 0 5Ms div to 5s div 5 and X Y 1 2 5 sequence Fine adjustment in all 19 ranges Magnification 10 times 5 PULL x 10 MAG Linearity Better than 3 2 zs div to 0 5s div Better than 5 0 5 s div to 1 s d...

Page 4: ...tation Trace angle is adjustable by panel surface ad justor Power Requirements Power supply voltage 100 120 220 240V 10 50 60 Hz Power consumption 47W Dimensions Width 260 mm 277 mm Height 190 mm 204...

Page 5: ...CONTROLS ON PANELS FRONT PANEL Fig 1 REAR PANEL Fig 2 5...

Page 6: ...Dual trace operation in which sweep is chopped at approximate 200 kHz rate and switched between Channel 1 and Channel 2 traces Recommended for sweep times of 0 5 s div to 1 ms div ADD The waveforms f...

Page 7: ...tive going sync pulse 19 LEVEL PULL AUTO Sync level adjustment determines points on waveform slope where sweep starts equals most negative point of triggering and equals most positive point of trigger...

Page 8: ...40V be sure to use a 0 5A fuse 29 POWER CONNECTOR For connection of the supplied AC power cord 30 Z AXIS INPUT Intensity modulation terminal T T L logic compatible high logic increase brightness low l...

Page 9: ...djust VOLTS DIV 4 7 and VARIABLE 5 8 so that the pulse amplitude is set to the 0 and 100 scales 2 Turn SWEEP TIME DIV 21 to magnify the rising portion of the waveform as large as possible VARIABLE 22...

Page 10: ...ative going waveform In this case the output waveform is shifted with respect to the leading edge of the Fig 5 Waveforms in divide by two circuit reference frequency pulse by a time interval cor respo...

Page 11: ...of the waveforms change depending upon the input or more mode of opera tion Ffg 8 shows a typical digital circuit and identifies several of the points at which waveform measure ment are appropriate T...

Page 12: ...uch cases it is recommended that the sync re main unchanged while the sweep speed or X 1 0 Fig 8 Typical digital circuit using several time related waveforms MAG used be to expaned the waveform displa...

Page 13: ...e amplifier unit may be measured for distortion with this oscilloscope This type of measurement is especially variable when the slope of a waveform must be faithfully reproduced by an amplifier Fig 10...

Page 14: ...ns Fig 11 shows a burst circuit The basic settings are identical to those of in Fig 5 Waveform EA is the reference waveform and is applied to CH1 input All other waveforms are sampled at CH2 and compa...

Page 15: ...Y P E T O H O R I Z O N T A L A M P L I F I E R T O V I D E O O U T P U T A M P L I F I E R N O T E NO E L E C T R I C A L C O N N E C T I O N P L A C E C L I P ON I N S U L A T I O N O F P L A T E C...

Page 16: ...ut jack The VARIABLE control is adjusted as required to provide a complete cycle of the input waveform displayed on 8 div horizontally A waveform height of 2 div is used The 8 div display represents 3...

Page 17: ...abled to see the VITS The transmitted VITS is precision sequence of a specific frequences amplitudes and waveshapes as shown in Fig 16 Television networks use the precision signals for adjustment and...

Page 18: ...s on either side This suggests an IF trap is detuned into the passband chopping out frequencies about 2 MHz below the picture carrier frequency Switch to another channel carrying VITS If the same thin...

Page 19: ...ection moving the trace to the left until the expanded VITS information appears as shown in Fig 16 Because of the low repetition rate and the high sweep speed combination the brightness level of the s...

Page 20: ...he peak to peak voltage readings on the oscilloscope with those shown on the waveform pictures Any abnormal readings should be followed by additional readings in the suspected circuits until the troub...

Page 21: ...controls for the desired brightness and best focus 15 To view a specific portion of the waveform such as the color burst pull outward on the POSITION control for X10 magnification Rotate the same con...

Page 22: ...tal gain controls for a display similar to that shown in Fig 23A 5 Set the marker generator precisely to 10 7 MHz The marker pip should be in the center of the bandpass 6 Align the IF amplifiers accor...

Page 23: ...e waveform A sine wave input is applied to the audio circuit being tested The same sine wave input is applied to the vertical input of the oscilloscope and the output of the tested circuit is applied...

Page 24: ...ces a circular oscilloscope pattern Phase shift of less or more than 90 produces an elliptical oscilloscope pattern The amount of phase shift can be calculated from the oscilloscope trace as shown in...

Page 25: ...can evaluate the input and output quality of a signal of many frequencies the harmoniques of the square wave which is what the amplifier sees when amplifying complex wave forms of musical instruments...

Page 26: ...ase a deliberate non linear circuit such as a clipper network 3 The third is delay or phase distortion which is distortion produced by a shift in phase between one or more components of a complex wave...

Page 27: ...lts are opposite in the two cases because of the difference in polarity of the phase angle in the two cases that can be checked through algebraic addition of components Fig 36 indicates low frequency...

Page 28: ...the square wave indicates a transient oscillation in a relatively high Q network in the amplifier circuit In this case the sudden transition in the square wave potential from a sharply rising relativ...

Page 29: ...tuated fundamental C High frequency loss No phase shift D Low frequency phase shift E Low frequency loss and phase shift F High frequency loss and low frequency phase shift G High frequency loss and p...

Page 30: ...or EXT TRIG is up to 50V DC AC peak and the input to Z AXIS is up to 50V DC AC peak 4 Do not increase the brightness of the CRT unnecessarily 5 Do not leave the oscilloscope for a long period with a b...

Page 31: ...adjustments use a well insulated screw driver 3 Before marking adjustments be sure to turn on the power and wait until the unit is stabilized 4 For adjustment follow the procedures described below 5...

Page 32: ...the trace comes to the position of the line in CH2 NOR mode when the CH2 polarity is inverted Horizontal Position Adjustment 1 To adjust the horizontal position under the nor mal sweep condition set t...

Page 33: ...SCHEMATIC DIAGRAM CS 1566A 33...

Page 34: ...A p r o d u c t of T R I O K E N W O O D C O R P O R A T I O N 17 5 2 chome Shibuya Shibuya ku Tokyo 150 Japan 2 5 3 1 2 PRINTED IN J A P A N B50 2918 00 G...

Reviews: