EMC Characteristics
EEM
MC
C D
DIIR
REEC
CTTIIVVEE 22000044//110088//EEC
C
Component power supplies such as the CoolX series are not covered by the EMC
directive. It is not possible for any power supply manufacturer to guarantee
conformity of the final product to the EMC directive, because performance is
critically dependent on the final system configuration. System compliance with the
EMC directive is facilitated by this product complying with several of the
requirements as shown in
Table 1
‑
3
. Although CoolX units meet these requirements,
the CE mark does not cover this area.
A full EN60601-1-2 4
th
Edition test report is available on request. Contact AE Global
Services for details.
TTaabbllee 11‑‑33.. EMC characteristics of CoolX unit under load conditions
PPaarraam
meetteerr
C
Coonnddiittiioonnss//D
Deessccrriippttiioonnss
C
Crriitteerriiaa
Radiated emissions
EN55011, EN55022 and
FCC, Class B
[1]
Conducted emissions
EN55011, EN55022 and
FCC, Class B
Power line harmonics
EN61000-3-2, Class A
Voltage flicker
EN61000-3-3
ESD
EN61000-4-2, Level 4,
8 kV contact, 15 kV air
A
Radiated immunity
EN61000-4-3, Level 3,
10 V/m
A
Electrical fast transient
EN61000-4-4, Level 4, ±
4 kV
A
Surge immunity
EN61000-4-5, Level 4,
2 kV DM, 4 kV CM
B
Conducted RF immunity
EN61000-4-6, Level 2,
3 V
RMS
A
Power frequency magnetic
field
EN61000-4-8, Level 4,
30 A/m
A
1
Some applications might require an external ferrite on cabling to meet Class B radiated
EMI. Contact AE Global Services for details.
Radiated EMI should be tested in a system environment. Radiated EMI performance
in a system will vary significantly from a standalone power supply because the
system enclosure provides additional shielding.
Advanced Energy
®
CoolX
®
600 Modular Power Supply
57010208-00B
Safety and Product Compliance Guidelines
1‑5