F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary unit
selected in F8.
"FS" = Full scale in primary units.
FS
FS + 2%
√
FS + 1d
FS + 9d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
√
2
F9
Display
Divisions
Determines the desired weight increments. Value should be consis-
tent with legal requirements.
1
√
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
√
0.0
0.00 0.000
0.0000 00
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the
ZERO
key one level begins
the procedure.
Press
ZERO
key
to begin sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the
ZERO
key one level be-
gins the procedure.
Press
ZERO
key
to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Press the
ZERO
key twice to exe-
cute.
TI-500E-SS (4-20) only
NAME/CODE DESCRIPTION CODE/VALUE
F23
Fine Tune
4-20 mA
Actuates the function that allows you to fine-tune the 4-20 mA analog
output. Pressing the
ZERO
key to scroll down one level begins the
sequence.
Press the
ZERO
key
Page 4-2
2
Also appears on newer TI-500E and TI-500E-SS indicators as well as older units with optional 4-20 mA analog
output.