3.2.5 EXITING THE SETUP MENU – ABS ENCLOSURE
1
. Power off the indicator.
2
. On the rear panel, move the Setup/Calibration Switch back to its original position.
3. Power on the indicator. The display will go through a digit check, then settle into Normal
Operating mode. All front panel keys will now return to their normal mode of operation.
3.2.6 EXITING THE SETUP MENU – STAINLESS STEEL ENCLOSURE
1. Power off the indicator.
2. Move the slide switch on the rear cover back to the left.
3. Power on the indicator. The display will go through a digit check, then settle into Normal
Operating mode. All front panel keys will now return to their normal mode of operation.
Note:
If your indicator is an older model and does not have the slide switch on the rear
cover, follow these instructions instead:
TI-500-ESS (without slide switch)
1
. Remove the rear cover and locate jumper JP2.
. Position the shunt block as shown at right.
JP2
2
Note:
On certain units, the shunt block position will be exactly the opposite.
3. Power on the indicator. The display will go through a digit check, then settle into Normal
Operating mode. All front panel keys will now return to their normal mode of operation.
TI-500-SSB (without slide switch)
1.
emove the rear cover and locate the two-position dipswitch.
R
2.
lace switch #1 to the ON position.
P
N
ote:
On certain units, the dipswitch position will be exactly the opposite.
3. Power on the indicator. The display will go through a digit check, then settle into Normal
Operating mode. All front panel keys will now return to their normal mode of operation.
3.3 USER (“A”) MENU
3.3.1 ENTERING THE USER MENU
1. Enter the Setup (“F”) menu by following the directions in Section 3.2.1 or 3.2.2.
2. Use the right or left directional keys shown in Figure 3-3 to move right or left in the
Setup (“F”) menu until the indicator shows ” A 1”.
Note:
On certain older units, the User (“A”) Menu is independent from the Setup (“F”) Menu.
To enter the User Menu on these units, first exit the Setup Menu Mode. Turn the unit off,
then press and hold the LB/KG key while powering the unit back on. When the screen
shows ” A 1” you may release the LB/KG key.
Page 3-4