NAME/CODE DESCRIPTION CODE/VALUE
F24
Piece Count
Enable
Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count
0
√
1
SUB-MENU TITLE
SELECTIONS
F1
Graduations
6,000 8,000
10,000 12,000
20,000 30,000
40,000 50,000
F3
Zero
Tracking
Band (SAZSM)
0d
5d
F5
Motion Band
3d 5d
10d 15 d
20d 30d
40d 50d
F6
Digital
Filter
1
2 4
F24
Piece
Count
1
Table 4-1: Invalid Setup Menu selections for commercial applications
4.2 SETUP MENU PROCEDURES
This section provides instructions for all of the Setup Menu procedures except F16 through F20.
These can be found in Chapter 6.
4.2.1 Fine-tune 4-20 mA output (F23)
1. While in the Setup Menu mode, scroll to "F 23", then scroll down once using the ZERO key to
enter Fine-tune menu. The indicator outputs 4 mA and displays a number.
2. While monitoring the voltage across R
L
(see Appendix E), use the right (PRINT) or left
(TARE) keys to change the displayed value until the measured voltage is exactly 1 VDC.
3. Press the SET (Net/Gross) key to save. The indicator outputs 20 mA and displays another
number.
4. While monitoring the voltage across R
L
, use the right (PRINT) or left (TARE) keys to change
the displayed value until the measured voltage is exactly 5 VDC.
5. Press the SET (Net/Gross) key to save and revert back to F23.
Page 4-3
3
Won’t appear on older units.