Neptune 6500
22
F15
Setpoint 2
delay time
Allows user to set time delay in 1/10s of a second to cause SP 2 to
alternate high and low when SP2 condition is met. A 0 (zero) value is
always on.
0
ÖÖ
Enter value
F16
Zero
Calibration
Places indicator into the zero calibration routine. Use the down arrow
key to begin the procedure.
Press down arrow
key to begin
sequence
F17
Span
Calibration
Places indicator into the span calibration routine. Use the down arrow
key to begin the procedure.
Press down arrow
key to begin
sequence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed. Use
the down arrow key to begin the procedure.
Press down arrow
key to begin
sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Use the down arrow key to begin the procedure.
Press down arrow
key to begin
sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Use the down arrow key to begin the procedure.
Press down arrow
key to begin
sequence
F21
Factory Reset
This sub-menu will reset all parameters in the
“
F
”
and
“
A
”
menu to the
default settings. USE WITH CAUTION!
Press down arrow
key to begin
sequence.
F23
Scale Capacity
Allows user to set the capacity of the scale.
2.5, 5, 10, 25, 50,
100, 250, 500,
1000