6. If the calibration was not successful, one of the error messages below will appear. Take
the indicated action to correct the problem, then perform a new calibration.
"Err0" - The calibration test weight or the adjusted keyed-in weight is larger than the full
capacity of the scale. Change the calibration test weight or check the input data.
"Err1" - The calibration test weight or the adjusted keyed-in weight is smaller than 1%
of the full capacity of the scale. Change the calibration test weight or check the input
data.
"Err2" - The internal resolution of the scale is not high enough to accept the calibration
value. Select a larger parameter for the Span Gain (F2). SEE APPENDIX C FOR
MORE INFORMATION.
5.4
VIEW CALIBRATION VALUES (F18)
Note: The values displayed in this procedure are valid only after a successful calibration
has been performed using F16 and F17.
1.
While in the Setup mode, scroll to "F 18", then scroll down once using the DOWN
ARROW (2) key to enter View calibration menu.
2.
The display will momentarily show "CAL 0" followed by a value. This value is the zero
calibration value and should be recorded in the table below.
3.
Press DOWN ARROW (2) again. Scale will momentarily show "CAL 1" followed by
another value. This value is the span calibration value and should also be recorded in
the table below. Press the up arrow key to return to upper level (F18).
INDICATOR
ZERO CALIBRATION VALUE
SPAN CALIBRATION VALUE
S/N:
Table 6-1: Calibration Value Table
5.5
KEY-IN ZERO CALIBRATION VALUE (F19)
Note: This procedure is intended for emergency use only in the case of non-volatile
memory loss. A valid zero calibration value, obtained from a successful F16
calibration procedure, must be used.
1.
While in the Setup mode, scroll using the right or left arrow keys to "F 19", then scroll
down once using the DOWN ARROW (2) key.
2.
The display will momentarily show "CAL 0", followed by a flashing zero. Use the four
directional keys (shown in Figure 6-1) to adjust the displayed value to the zero
calibration value.
3.
After setting the exact value, press the ENT key to save the value.
4.
The display will show "E CAL 0" momentarily, then revert back up to F19.