154
TMX428 TMX428CP & TMX58 Service Manual
317-160-406 Rev 4
May 2021
Cleaning: Manual
Step 1
–
Turn the main power switch to OFF and disconnect the treadmill
from its power outlet.
Step 2
–
Use 20 mL (0.68 oz) of an alkaline cleaner or similar mild, non-
abrasive detergent (ex. Spray 409) on a soft, clean, lint-free 25 x
25 cm (9.8 x 9.8 in) size cloth / paper towel and manually clean
item 1, 2, & 6 (if applicable). Reference Image 1 for corresponding
items.
Note:
Do not spray cleaning chemical directly on the device
Step 3
–
Use an appropriately sized soft-bristled brush (e.g. standard
toothbrush) dampened with 5 mL (0.17 oz) of mild soapy tap water
(or similar mild, non-abrasive detergent) to clean item 3 and hard
to reach spots. Reference Image 1 for corresponding items.
Step 4
–
Wipe all surfaces 1-6 (reference Image 1) with a soft, clean, lint-
free 25 x 25 cm (9.8 x 9.8 in) size cloth / paper towel dampened
with 20 mL (0.68 oz) (not wet) of lukewarm tap water 27°C to
44°C (80°F to 111°F) for a minimum of 30 seconds.
Step 5
–
Dispose of all cleaning materials according to your institutions
established procedure.
Disinfection
If necessary, to disinfect the treadmill, follow these steps.
Step 1
–
Turn the main power switch to OFF and disconnect the treadmill
from its power outlet.
Step 2
–
Prepare 20 mL (0.68 oz) of 70% or greater isopropyl alcohol
disinfectant solution according to the manufacturer’s directions.
Step 3
–
Use a soft, clean, lint-free 25 x 25 cm (9.8 x 9.8 in) size cloth /
paper towel or gauze that is saturated with 20 mL (0.68 oz) of the
disinfection solution and manually disinfect all contaminated
surfaces of the device for a minimum of 15 minutes.
Step 4
–
Use an appropriately sized soft-bristled brush (e.g. standard
toothbrush) dampened with 5 mL (0.17 oz) of the disinfectant
solution to clean item 3 and hard to reach spots. (reference Image 1
for corresponding items) Surface must remain wet for a minimum
of 15 minutes.
Step 5
–
Wipe all contaminated surfaces 1-6 (reference Image 1) with a soft,
clean, lint-free 25 x 25 cm (9.8 x 9.8 in) size cloth / paper towel or
gauze dampened with 20 mL (0.68 oz) of purified water 27°C to
44°C (80°F to 111°F) for a minimum of 30 seconds.
Step 6
–
Dispose of all cleaning materials according to your institutions
established procedure.
Drying
Dry the device with a soft, clean, lint-free cloth/paper towel 25 x 25 cm (9.8
x 9.8 in).
Maintenance,
Inspection and
Testing
Visually examine each device for cleanliness. If visible soil remains, repeat
cleaning procedure until the device is thoroughly clean.
Packaging
Not applicable
Sterilization
Not applicable