Open the cover of the CureView.
Push the spring behind the probe block
backward.
Push back the base of the probe block at
the notch and lift the probe block out.
Fix the nut below the controller.
Replace the controller with a new one.
Take care not to switch the controllers or
place them at the wrong locations. Do
not exert force on the cables connecting
the probe to the controller.
Lift the controller and guide the probe
block out through the back of the
CureView.
10
13
11
12
15
14
Reconnect the connectors. Take care
the TQC-BUS connector(s) is/are placed
correctly.
Raise the controller holder plate and
tighten the allen keys (3mm) on each
side.
Tilt and slide in the rear cable cover back
in place. Fix the 2 allen keys to hold it in
place.
If the cables prevent a smooth
positioning. Use the allen key to gently
push the cables down. To allow to slide
the cable cover back in place.