10.4 Exhaust system
The CureView has a self-powered fume exhaust system mounted on the
rear. This system is created to allow fumes emanating from the test panel to
be expelled safely. The CureView may be connected to an already existing
ventilation system when this is a system without forced convection. If
the ventilation system has forced convention care should be taken not to
disturb the flow of air in the CureView.
10.5 Installation procedure
Verify that above criteria are met. If above criteria are met proceed as
followed:
Remove all packaging material from the CureView and store it
for possible later use.
Lift the CureView onto the the worktable by lifting it with 2
persons. Each grabing the machine from the side.
Connect the exhaust.
Connect the power supply.
Place the USB cable.
Install the supplied laptop.
Connect the USB cable to the laptop
Power up the TQC CureView.
Power up the laptop.
Verify that 2 beeps are sounded from the gradient oven.
Start the Gradient Oven Control software.
No error message should be shown.
The TQC CureView is designed to operate in general lab conditions on a
sturdy worktable. In order to create the safest work environment possible
below requirements should met.
10.1 Climate
It is recommended to operate the TQC CureView in a temperature stable
environment. Due to the general tests performed with the TQC CureView a
climate controlled environment is advised having 20-25˚C / 68-77˚F and
40 – 60 %rH. Other ranges of temperature and humidity can limit the
minimal programmable temperature of the machine. For example if the
ambient temperature is 35˚C / 95˚F the minimal programmable temperature
is approximately 10˚C / 50˚F higher.
10.2 Power Supply
The TQC CureView is designed to operate at 220 - 240V / 50/60Hz power
supply, this due to the power consumption of 3500W. For the 230V /60Hz
equipped countries the 3500W is together with the laptop the maximum
power that can be drawn from a fuse. The CureView should be run on a
dedicated fuse. The TQC CureView cannot be connected to 110V power
supply. When the TQC CureView is intended for use in 110V countries the
CureView needs to be connected to Split phase power supply. Please
consort a suitable trained technician to verify the power supply.
There should be a second power plug available to
connect the laptop.
10.3 Worktable
A suitable and sturdy table needs to be selected with sufficient depth and
clearance above the table to operate the TQC CureView. No power supplies,
sockets or such may be placed behind the CureView that could block or
be affected by the heat exhaust in the rear of the CureView. The worktable
should be able to support 40 kg / 88.18 lbs of load and allow for placement
of the CureView with following clearances. The height of the table should be
selected such that ergonomic and safe operation is the machine is possible.
Please consult local guidelines and regulations.
10 INSTALLATION CRITERIA
1
3
5
7
9
11
2
4
6
8
10
12
> 600 mm
< 150 mm
Exhaust
fumes