17.1.2
Slope of a pH Electrode
As mentioned above, a pH electrode produces 0 mV output at around 7.00 pH. As the pH goes up,
an “ideal” pH electrode produces -59mV/pH unit at 25
O
C As the pH goes down, an ideal pH
electrode pr59mV/pH unit. In practice, pH electrodes usually produce slightly less than
this. The output of a pH electrode is expressed as a percentage of an ideal electrode. For example,
an ideal electrode that produces 59mV/pH unit has “100% Slope”. An electrode that produces
50.15mV/pH unit has “85% Slope” (see Figure 17-2).
-600
-400
-200
0
200
400
600
0
7
14
mV
85% Slope at 25 oC
(50.15mV/pH)
100% Slope at 25 oC
(59mV/pH)
Response of pH Electrode, as a Function of Slope
pH
Figure 17-2
17.1.3
Temperature Compensation
The slope of a pH electrode is affected by temperature. This effect is compensated for either by
using an Automatic Temperature Compensation (ATC) probe or by entering the sample temperature
manually. Figure 17-3 shows the slope of a pH electrode at various temperatures.
-600
-400
-200
0
200
400
600
0
7
14
mV
Electrode
Potential (mV) at
0 oC (54mV/pH)
Electrode
Potential (mV) at
50 oC (64mV/pH)
Electrode
Potential (mV) at
100 oC
(74mV/pH)
pH Electrode Response, as a Function of Temperature
pH
Figure 17-3
17.2 Guard Connector
In some circumstances, the Specific Ion, pH or mV readings may become unstable. This may be
due to static charge in the sample vessel, or electrical noise from nearby electrical equipment. In
these cases, a solution guard may eliminate the problem.
A solution earth rod is available from TPS (part no 121360). This connects directly to the
Guard
socket. Alternatively, run a wire from the
Guard
socket to a stainless steel fitting in contact with
the sample.