37
12. Recorder Port
This section is applicable if the optional Recorder port is fitted.
The optional Recorder Port can be used to send any one Channel of Specific Ion, pH, mV or
Relative mV data to a chart recorder or other analogue logging device.
The Recorder Port is not activated until it has been configured, as per sections 12.1, 12.2 or 12.3. If
the readout mode of the selected Channel is changed, the Recorder Port is de-activated to prevent
potentially misleading outputs.
12.1 Recorder Port Configuration in Specific Ion Mode
1.
Select the Ports Set-up menu (
→
→
F4:Setup
F4:Setup
→
→
F1:Ports
F1:Ports
).
2.
If the both the Recorder and the RS232 port options are fitted, then select
F2:Recorder
F2:Recorder
.
If the Recorder Port option
only
is fitted, then the
900-I3
will jump directly to the Channel
Selection screen…
RECORDER F1:Channel 1 F2:Channel 2
RECORDER F1:Channel 1 F2:Channel 2
F3:Channel 3 >F4:OFF
F3:Channel 3 >F4:OFF
The arrow indicates the current selection.
Press
,
or
to select Channel 1, 2 or 3.
Press
to switch the Recorder Port OFF.
3.
After selecting Channel 1, 2 or 3, the first Recorder Port configuration screen is displayed…
Select Option >F1:Ion F2:mV
Select Option >F1:Ion F2:mV
The arrow indicates the current selection.
Press
to set the Recorder output to be directly proportional to the concentration readout on
the
900-I3
display.
Press
to set the Recorder output to be proportional to the millivolt signal being produced
by the Ion Selective Electrode. This millivolt signal is temperature compensated for the
Recorder output.