11.5 Commands
The following commands can be sent from a PC to the
900-I3
. Note that <cr> denotes carriage
return and <lf> denotes a line feed.
Action
Command
Notes
Request current data
?D<cr>
Returns current data from all
Specific Ion/pH/mV Channels that
are switched on, plus Temperature,
date and time from the
900-I3
. The
log number returned is set to Zero.
Request logged data
?R<cr>
Returns all logged records from the
900-I3
memory. The data ends with
the message
ENDS
<cr>
Erase logged data
?E<cr>
Erases all logged records from the
900-I3
memory. Returns the
message
ERASED
<cr> to confirm
that the records have been erased.
Request status
information
?S<cr>
Returns the model name, firmware
version number, instrument serial
number and number of logged
readings in memory, for example…
900I3
s
V3.0
s
R1234
s
1200
<cr>,
where
s
are spaces. Note that the
number of logged readings is right-
justified.
Request GLP
information
?G<cr>
Returns all calibration GLP
information, plus the instrument
model, serial number and current
date (see section 11.7 for data
format and handshaking).