40
12.4 Recorder Port Specifications
Mode
Display Range
Output Range
Examples
(Reading = mV Out)
Specific Ion
Full auto-ranging
concentration readout, with
user-set resolution.
2000 mV at user-selectable
Full Scale ( section 12.1).
0 mV at approx. 2 decades
of concentration below Full
Scale.
If Full Scale is set to
1000ppM, then…
10 ppM
≈
0 mV
1000 ppM = 2000 mV
pH
0 to 14.000 pH
0 to 14.00 pH
0 to 14.0 pH
0 to 2000 mV
(for all resolutions)
7.00 pH = 1000 mV
Absolute mV,
ISE mV
-1500 to +1500 mV
0 to 2000 mV
0.0 mV = 1000 mV
+750 mV = 1500 mV
-600 mV = 600 mV
Relative mV
-1500 to +1500 mVR
0 to 2000 mV
0.0 mVR = 1000 mV
+750 mVR = 1500 mV
-600 mVR = 600 mV
Output Impedance
: Approximately 1000 Ohms
Resolution
: Approximately 2 mV
12.5 Recorder Port Connections
1
5
6
9
Pin No
Connection
6
Recorder Output Signal
7
Recorder Output Common