5. STORING/SAVING DATA
It is possible to save complete combustion readings, temperature and single
pressure readings as follows:-
1.
Have the 717R analyser set to the relevant screen for the readings that
you wish to save (i.e. Any of the combustion screens (see 4.1) for
combustion readings or pressure screen (see 4.3.1) for single pressure
reading OR have just completed a Tightness or CO Build Up Test)
2. Press the
Store/Save Key
once
3.
‘Addr’
will be displayed and a location number (0
to 99) will be flashing on the screen.
4.
Select the required address location that you wish
to save the data to by pressing
the
Up and Down Arrow Keys
**Please Note:- Choosing the same Address Location on which previously
Stored Data is being held will
OVERWRITE
the previous readings with the New
Stored Data. This will be the case regardless of the particular readings you are
attempting to save (i.e. should you have Combustion Readings stored at Address
1 and you Save Temperature Readings to Address 1 then the Previous
Combustion Readings will be overwritten with the New Temperature Readings)
and the previously Stored Readings will not be retrievable.
5. Press the
Enter Key
once and
the readings will be stored to the
location that was chosen.
6. The 717R will then return to the Last Screen visited.
The information that you just stored can be either be Reviewed on the screen
(see 7) or Printed to a compatible IR Printer (see 8)