19. Once the countdown is finished the 717R will
display P1, P2 and Differential (P2-P1) and give
you the option to “Pass” or “Fail” depending on
whether the readings are within the allowable
tolerance or not.
20.
Choosing “Fail” by pressing the
Down
Arrow Key
will stop the test and allow you to, if
required, store the readings so far.
21.
Choosing “Pass” by pressing the
Up
Arrow Key
will take you to the Finish screen
22. From the Finish Screen you can store/save (see
section 5) or print (see section 7) the
readings/test
4.5 Function 5: - CO Build Up Test
Remember:-
It was the
Function Key
that got you here!!!
The CO room test function enables the 716 to monitor and log ambient CO levels
in a room or office space at 1 minute interval up to a total of 50 minutes. This data
can be retrieved later via the optional infrared printer.
1. Begin with the 717R in a fresh clean air
environment
2. Press the
Enter Key
start to the CO Room
Test
3. The 717R will start and record the CO level every
60 seconds and count up to 50 on the screen
4. At any time you can stop the test by pressing the
Enter Key
5. At the end of the test you can store/save (see
section 5) or print (see section 7) the readings