Reassembly of Meter
Reinstalling Packing Capsule & Front/Rear Covers
1) Reinstall cover O-rings on the front and rear of meter housing.
2) Reinsert packing capsule into front cover of meter with retaining plate and tighten screws.
3) When attaching front cover, align packing capsule blade with the slot on the blocking rotor sleeve. Tighten all
front cover screws.
4) Attach rear cover and tighten all screws.
Reinstalling Counter Support & Adjustor Drive Shaft
1) Position the counter support on the front side of meter and tighten all screws.
2) Reinsert adjustor drive shaft into the support housing and assure that adjustor face gear teeth mesh with the
packing capsule gear. Do not fit the face gear and packing gear too tight!
3) Slide down the adjustor mounting bracket and tighten the screws.
1
2
3
4
1
2
3