15
SYSTEM DIAGRAM
Photography screen (Chin-rest adjustment): Advance mode
When you set "ON" for the "Advance" mode on the screen where you adjust the chin-rest as watching
the anterior segment observation image and the image of the anterior segment stereoscopic camera,
this screen appears.
Small pupil diaphragm
selector button
: Displays the small pupil diaphragm status (ON/OFF). You can change
the ON/OFF status of the small pupil diaphragm. When the pupil diame-
ter of the patient is small (pupil diameter is approx.
4.0mm or less), set
it to ON and take a picture.
When the color fundus photography is done (excluding the color fundus
photography when taking the fundus tomogram), the automatic small
pupil (diaphragm) function can be set. While this function is set, the
small pupil diaphragm is automatically set to ON and the instrument
takes a color picture of fundus if the pupil diameter is small (pupil diame-
ter is approx.
4.0mm or less). After taking a picture, it is set to OFF.
When you set this function, in the color fundus photography mode, it is
not necessary to set the small pupil diaphragm to ON with this button
even if the pupil diameter is small.
Color fundus photography
ON/OFF button
: You can turn ON/OFF the color fundus photography. When the instru-
ment without the fundus photography function is used and when the
color fundus photography function is OFF, this button is not displayed.
1234ABCD
Diopter compensation
lens selector button
[MENU] button
Small pupil diaphragm selector button
Cataract mode button
Color fundus photography ON/OFF button