SOL-96.110.061 / 031221
TOPAZ Manganese
Troubleshooting
58
43
Division by zero - Absorbance =
-999998.0
– Check that the optical fibers are well
connected to their holder on the door.
44
Value out of range
– Check that the analyzer is well
calibrated, check standard solution
concentration by a manual measure
cycle.
45
Division by zero - Absorbance =
-999998.0
– Check that the optical fibers are well
connected to their holder on the door.
46
Incorrect projector power
– Contact customer service
47
Threshold 1 exceeded
– Check process
– Check programmed value for
threshold 1
48
Threshold 2 exceeded
– Check process
– Check programmed value for
threshold 2
49
Optical setting incorrect
(analyzer cannot adjust the
optical gain)
– Check that the optical fibers are well
connected to their holder on the door.
– Check that the photometer is not dirty.
– Check that the analyzer is supplied
with water.
– Check that the stirrer turns.
No.
Text shown on display
Corrective action