SOL-96.110.061 / 031221
TOPAZ Manganese
Maintenance
34
6.3. Replacing Reagents
WARNING
Multiple hazards
For safe handling of the reagents, you must read and understand
the Material Safety Data Sheets (MSDS). Wear appropriate
protective equipment.
Note:
• It is recommended to replace the reagents when there is
approximately 2–3 cm of reagent left in each bottle. Do not use
up the reagents completely!
• Keep the old reagents until you are sure that the instrument
measures correctly with the new reagents.
• Do not reuse the reagent bottles as this poses a risk of
contamination or dilution errors.