TH2683A Operation Manual
Ver.1.3
42
Where, <value> is any real value among the charge time range (0
‐
999.0s) of the instrument
(correct to 0.1s).
For example: WrtCmd(“FUNCtion:CTTM 12.5”); set the charge time of the instrument as 12.5s.
Query syntax: FUNCtion: CTTMe?
Return Format: <value><NL^END>
The :FUNCtion:CTTMe? query returns to thecharge time of the instrument.
The :WTIMe command set the wait time of the instrument.
The FUNCtion:WTIMe? query returns to the current wait time of the instrument.
Command syntax: FUNCtion: WTIMe <value>
Where, <value> is any real value among the wait time range (0
‐
999.0s) of the instrument (correct
to 0.1s).
For example: WrtCmd(“FUNCtion:WTIme 12.5”); set the wait time of the instrument as 12.5s.
Query syntax: FUNCtion: WTIMe?
Return Format: <value><NL^END>
The :FUNCtion:WTIMe? query returns to the wait time of the instrument.
The :MTIMe command set the measurement time of the instrument.
The FUNCtion:MTIMe? query returns to the current measurement time of the instrument.
Command syntax: FUNCtion: CTIMe <value>
Where, <value> is any real value among the measurement time range (0
‐
999.0s) of the
instrument (correct to 0.1s).
For example: WrtCmd(“FUNCtion:MTIMe 12.5”); set the measurement time of the instrument as
12.5s.
Query syntax: FUNCtion: MTIMe?
Return Format: <value><NL^END>
The :FUNCtion:MTIMe? query returns to the measurement time of the instrument.
The :DTIMe command set the discharge time of the instrument.
The FUNCtion:DTIMe? query returns to the current discharge time of the instrument.
Command syntax: FUNCtion: DTIMe <value>
Where, <value> is any real value among the discharge time range (0
‐
999.0s) of the instrument
(correct to 0.1s).
For example: WrtCmd(“FUNCtion:DTIMe 12.5”); set the discharge time of the instrument as
12.5s.
Query syntax: FUNCtion: DTIMe?
Return Format: <value><NL^END>
The :FUNCtion:DTIMe? query returns to the discharge time of the instrument.
The :AVERage command sets the measurement average time of the instrument.