24 

2.

 

INSERT, DELETE, NEW, SAVE or select up or down by pressing the F1~F6. 

3.

 

Press the right arrow key to move the cursor to the test mode option, as shown in Figure 4-12, 

uses  the  F1~F5  keys  to  select  whether  the  test  mode  is  AC  withstand  voltage,  DC  withstand 

voltage, insulation resistance, pause mode or open short circuit detection OSC. 

4.

 

Move the cursor to the test parameter in the corresponding mode by using the  arrow keys and 

change the parameter setting by using the F1~F6 keys or numeric keys. 

                           

 

Figure 4-11 Meas Setup 

Test  procedures:  current  setting  step  number  /  total  number  of  test  steps.  The  change  of  the  test 

steps are as shown in the table below: 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Key        Function 

Explanation 

F1 

INSERT 

Add  a  new  test  item.  The  current  item  and  subsequent  items 

will move one step backward. 

F2 

DELETE 

Delete  the  current  test  item.  The  subsequent  item  will  move 

one step forward. 

F3 

NEW 

Create  a  blank  test  program  (PROG).  Please  remember  to 

save the test program after finishing writing. 

F4 

SAVE 

Save the current test file 

F5 

Up 

Move to the display screen of the previous step of the current 

step. 

F6 

Down 

Move to the display screen of the next step of the current step. 

Table 4-3 

F1















4-2


1















4-2


co

F2

co

F3

co

F4

  F1

F2

co

F3

co

F4

co

F5 















4-2


1















4-2

F6 

co

Summary of Contents for TH9110

Page 1: ...1 TH9110 TH9110A AC DC Withstanding Voltage Resistance Tester V1 5 2018 07 Operation Manual Tonghui Electronic Co Ltd www tonghui com cn...

Page 2: ...Basic operation 15 4 1 Interface structure overview 15 4 2 SYSTEM Setup 16 4 2 1 System Test Parameter Settings 16 4 2 2 System Environment Parameters Setting 21 4 2 3 System Communication Setup 23 4...

Page 3: ...S232C Interface Instruction 45 5 1 2 GPIB Interface Instruction 51 5 2 Serial Port Commands Instruction 54 5 2 1 SCPI Commands 54 5 2 2 DISPlay Subsystem Commands 55 5 2 3 FUNCtion Subsystem Commands...

Page 4: ...enable the device to work with various automatic testing systems that require high security and reliability The tester provides multiple test functions typical ones High Voltage Floating Output desig...

Page 5: ...Operating Temperature 20 8 Humidity 80 RH Storage Ambient Temperature 10 55 Humidity 90 RH 1 2 3 Preheating Warm up time after power on 20 minutes 1 2 4 Precautions 1 Please do not locate and use the...

Page 6: ...otectors of the tester may not be able to render their protective functions properly Do not turn OFF the power switch when the tester is delivering its test voltage you may do this only in case of eme...

Page 7: ...Electromagnetic Compatibility Power Transient Sensitivity refer to requirements of GB6833 4 Conductance Sensitivity refer to requirements of GB6833 6 Radiative Interference refer to requirements of GB...

Page 8: ...resulting in extreme danger anyone who touches such tester chassis will result in electric shock accident therefore please make sure that the protective grounding terminal is connected to the safety g...

Page 9: ...to discharge after the output is cut off There is a danger of electric shock during discharge do not touch anything that could cause an electric shock As soon as the output is cut off the tester s dis...

Page 10: ...he high voltage output terminal HV1 to HV2 or RTN LOW the entire housing of the tester must be reliably connected to the ground 11 Do not connect external voltage to the test terminal Do not apply a v...

Page 11: ...ristics etc If different type of fuse is used or short circuit then the tester may be damaged 6 AC Power source The AC input power used by this tester is 100V 240VAC frequency is 47Hz 63Hz If the powe...

Page 12: ...rnally connect to USB storage 4 Brand and Model 5 PrtScn key Screenshot key it is used to capture the current screen image to the USB memory and USB need to be inserted into the front panel jack in ad...

Page 13: ...potential terminal of high voltage output this is the high voltage output terminal when DANGER light is on it means there is high voltage delivering touching is forbidden 13 HV2 Terminal RTN terminal...

Page 14: ...uter 7 LAN Interface LAN communication interface can realize the communication with the computer 8 USB Device interface Through USB DEVICE communication interface the computer can control the instrume...

Page 15: ...amme and the total steps is 1 AC means the AC withstanding voltage test interface AC parameter means that other parameters are AC withstanding voltage test parameters The third line in the interface i...

Page 16: ...lated system settings 3 Press key to move the cursor to the parameters you want to set Change the parameter settings with F1 F6 keys or numeric keys 4 If you need to input with the numeric keys press...

Page 17: ...an be set to continue the next test or press START to restart or press STOP first then press START to start the test PassHold 0 2 99 9S 0 5S Set the duration of the buzzer sound when in PASS StepHold...

Page 18: ...STOP key before pressing the START key to restart the test 4 2 1 3 Hardware Software Compensation Due to the output impedance of the test signal source the load effect that is the load change the outp...

Page 19: ...easured current can be expressed by the low current range the current range is automatically changed to the low current range Examples are shown in Figure 4 4 1 and Figure 4 4 2 and Figure 4 4 3 1 If...

Page 20: ...ting has three options namely OFF ON and FLOAT only TH9110 has high voltage floating output function OFF and ON means the function is turned off and on FLOAT means that the two high voltage output HV1...

Page 21: ...hus ensuring the safety of operator As shown in Figure 4 7 Figure 4 6 Figure 4 7 Relationship of HV1 HV2 Terminal to the Ground 4 2 2 System Environment Parameters Setting Operation Instructions 1 Pre...

Page 22: ...ON or OFF LCD Bri 1 10 5 Adjust the screen brightness Fail Vol OFF ON ON Set the FAIL beep to ON or OFF Date Set Set the year month and day as needed Time Set Set 24 hour time as needed Table 4 2 Key...

Page 23: ...be set and change the parameter setting by F1 F6 or numeric key System communication setup as shown in Figure 4 10 Figure 4 10 system communication setting As can be seen from the above figure the co...

Page 24: ...up Test procedures current setting step number total number of test steps The change of the test steps are as shown in the table below Key Function Explanation F1 INSERT Add a new test item The curren...

Page 25: ...imit 0 001mA Hi Limit 0 CLOSE CLOSE Set the low limit of the test current The low limit must be high limit or Close Arc Limit 1mA 20mA 0 CLOSE CLOSE Set the allowable AC arc current high limit CLOSE m...

Page 26: ...stand voltage testing Hi Limit 0 0001mA 25mA 500 0uA Set the high limit of the test current The voltage is 1 5kV the maximum current can reach 20mA the voltage is 1 5kV the maximum current can reach 2...

Page 27: ...age When Ramp Time is set to Close the voltage rise is completed during the first cycle of the test DwellTime 0 1 999 0s 0 CLOSE CLOSE Set the time needed for dwell the current high limit and low limi...

Page 28: ...ltage Close the ramp time the voltage ramp is completed during the first cycle of the test MeasTime 0 3 999 0s 0 Continue 3 0S Set the measurement time for the insulation resistance test 0 indicates t...

Page 29: ...5M 3mA 0 3 3mA 0 5M 4 5M 3 0M 15 0M 300uA 30 300uA 3 0M 15 0M 10 0M 45M 30uA 3 30uA 10 0M 45M 35 0M 450M 3uA 0 3 3uA 45M 0 45G 0 40G 4 5G 300nA 20 300nA 0 40G 4 9G 4 0G 50 0G Table 4 6 4 3 4 Pause Mod...

Page 30: ...ng the numeric keys or you can obtain a value as a standard value by sampling Table 4 7 Sampling Move the cursor to Sampling as shown in Figure 4 17 Input the standard value by using numeric key then...

Page 31: ...for details The upper one is the high voltage output voltage and the withstand voltage s unit is kilovolts kV The middle one is the test value the withstand voltage testing shows current unit is mill...

Page 32: ...erface Figure 4 18 3 IR Test Interface F1 4 2 1 4 2 co m F2 co m F3 co m F4 F1 F2 co m F3 co F4 co m F5 4 2 1 4 2 co m F6 co m F1 4 2 1 F2 co m F3 co m F4 F1 F2 co m F3 co F4 co m F5 4 2 1 F6 co m F1...

Page 33: ...arts to measure the leakage current of the test cable and display the current use the present test value as the zero value 3 After pressing F1 offset key the DANGER indicator flashes and the voltage o...

Page 34: ...After pressing START to start the test the third column shows the present voltage test value and the fourth column shows the actual measured current value or resistance value As shown in Figure 4 20 4...

Page 35: ...tage output 3 Finally connect the test cable of low potential RTN LOW or HV2 to the DUT first and then connect the test cable of high potential HV1 to the DUT 4 5 4 Test Procedures 4 5 4 1 AC DC IR Te...

Page 36: ...repare for the test Press the START key to start the test when this key is pressed there will be a high voltage output and the DANGER light will flash The first line of the large font shows the actual...

Page 37: ...s 350P 400P 88 and it is recommended to take 60 3 SHORT value the low limit is 450P 400P 112 the high limit is550P 400P 138 and it is recommended to take 125 4 6 Voltage Breakdown Test 4 6 1 Brief des...

Page 38: ...age value Meas Mode AC DC AC Measurement mode selection AC DC Ve 0 05 5kV 0 05kV Set voltage ending value Vs 0 05kV Termination Voltage 0 05kV Set voltage starting value Ramp Time 0 999 0 CLOSE CLOSE...

Page 39: ...File Storage Operation Instructions 1 Press FILE key to enter the file management interface as shown in Figure 4 24 2 Through the F1 key F2 key to switch between internal files and external files 3 Mo...

Page 40: ...esult output signal Signals can be divided into two broad categories internal output signals and external input signals The internal output can be divided into test result output and contact signal ou...

Page 41: ...stem generates an internal error 16 EOT Output When this signal is HIGH it means the test is in process when this signal is LOW it means the test is finished or in standby 17 EOS Output When this sign...

Page 42: ...42 Take the External Power Supply as an Example...

Page 43: ...43...

Page 44: ...nal trigger signal EXT_STRAT time needed from start to EOT signal is cleared 20mS T3 Time setup for the trigger delay T4 Time needed for each test steps STEP T5 10mS PASS_FAIL signals stable waiting t...

Page 45: ...d which can also be called the asynchronous serial communication standard which is used to realize data communication between a computer and a computer and between a computer and a peripheral device R...

Page 46: ...omputer The tester is connected to the computer as shown As can be seen from the above figure the pin definition of this tester is the same as that of the 9 core connector serial interface pin used by...

Page 47: ...do not receive the returned characters there may be factors 1 Serial port connection failure 2 Check that the RS232 port function is enabled 3 The tester is executing a bus command and is temporarily...

Page 48: ...main function can be used by the user to expand the communication function arbitrarily while other sub functions illustrate how to use the serial port for input and output of the string define PORT 0...

Page 49: ...PORT 256 c read_port PORT break if n break if c ps ps else printf nE10 Write Echo Error exit 1 send_port PORT n send command end symbol delay 2 while check_stat PORT 256 read_port PORT read string fr...

Page 50: ...ar read_port int port union REGS r r x dx port serial port r h ah 2 int14 function2 read character int86 0x14 r r if r h ah 128 if ah 7 be set mean trans error printf nE01 Serial port read error exit...

Page 51: ...board is an option The control command system is open and the user can use the computer operation interface provided by the tester or can program the system according to the control command to achieve...

Page 52: ...52 Figure5 1 GPIB Connector Pin Structure Diagram GPIB Cable Connection Method 1 Tester Back Connector...

Page 53: ...er contact functions T5 Basic talk function Talk only function Talking cancelled when MLA Do not support serial roll call L4 Basic listen function Listen cancelled when MTA No listen only function RL1...

Page 54: ...3 The data of the command is an ASCII string The default format of the system is integer or floating point number The default data unit does not appear in the commands 4 The command end must have an...

Page 55: ...e to Standby Home Page Character can query the current page Example Set the display page to measurement display Command Syntax DISP PAGE TEST Query syntax DISPlay PAGE Return format TEST Set measureme...

Page 56: ...STEP 1 AC VOLT 1000 Query message FUNC SOUR STEP 1 AC VOLT Return value 1000 FUNC SOURce STEP AC UPPC To set inquiry about the UPPER current for ACW test Syntax Command message FUNC SOUR STEP sn AC UP...

Page 57: ...for ACW test Syntax Command message FUNC SOUR STEP sn AC TTIM time Query message FUNC SOUR STEP sn AC TTIM Data time Data format float Data range 0 0 3 999 0s 0 is OFF Data accuracy 0 1s Data unit s...

Page 58: ...ge FUNC SOUR STEP 1 AC FTIM 1 Query message FUNC SOUR STEP 1 AC FTIM Return message 1 0 FUNC SOURce STEP AC ARC To set inquiry about the ARC upper current for ACW test Syntax Command message FUNC SOUR...

Page 59: ...Data sn Data format integer Data range 1 50 Data accuracy 1 Data voltage Data format float Data range 50 6000 Data accuracy 1 Data unit V Example Set the voltage for DCW test in STEP1 to 1000V Command...

Page 60: ...alue 1 000 FUNC SOURce STEP DC TTIM To set inquiry about the TEST time for DCW test Syntax Command message FUNC SOUR STEP sn DC TTIM time Query message FUNC SOUR STEP sn DC TTIM Data time Data format...

Page 61: ...TIM Return value 1 0 FUNC SOURce STEP DC WTIM To set inquiry about the wait time for DCW test Syntax Command message FUNC SOUR STEP sn DC WTIM time Query message FUNC SOUR STEP sn DC WTIM Data time Da...

Page 62: ...est in STEP1 to 1mA Command message FUNC SOUR STEP 1 DC RAMPARC 1 Query message FUNC SOUR STEP 1 DC RAMPARC Return message 1 0 FUNC SOURce STEP DC RAMPARC To set inquiry about the RAMP state for DCW t...

Page 63: ...imit 50 0G 0 is OFF Data accuracy 1 0 M Data unit M Example Set the upper resistance for IR test in STEP1 to 1 M Command message FUNC SOUR STEP 1 IR UPPR 1 Query message FUNC SOUR STEP 1 IR UPPR Retur...

Page 64: ...IR RTIM Data time Data format float Data range 0 999 0s 0 is OFF Data accuracy 0 1s Data unit s Example Set the rise time for IR test in STEP1 to 1s Command message FUNC SOUR STEP 1 IR RTIM 1 Query m...

Page 65: ...SOUR STEP sn PA MESSAge Data message Data format characters Example Set the PA message in STEP1 to HELLO Command message FUNC SOUR STEP 1 PA MESSAge HELLO Query message FUNC SOUR STEP 1 PA MESSAge Ret...

Page 66: ...Query message FUNC SOUR STEP sn OS SHOT Data rate Data format integer Data range 0 100 500 0 is OFF Data accuracy 10 Data unit Example Set the SHOT rate for OS test in STEP1 as 100 Command message FUN...

Page 67: ...mode Query message FUNC BREAK STEPMODE Data step mode Data format integer Data range 0 1 0 normal step 1 voltage step Example Set the step mode for breakdown test to 1 voltage step Command message FU...

Page 68: ...NC BREAK MODE Data mode Data format integer Data range 0 1 0 AC 1 DC Example Set the mode for breakdown test 1 DC Command message FUNC BREAK MODE 1 Query message FUNC BREAK MODE Return value 1 FUNC BR...

Page 69: ...ry message FUNC BREAK UPPC Data current Data format float Data range 0 0001 25mA DC 0 001 100mA AC Data accuracy 0 0001mA DC 0 001mA AC Data unit mA Example Set the upper current for breakdown test to...

Page 70: ...ue 1 0 FUNC BREAKdown RTIM To set inquiry about the RISE TIME for voltage breakdown test Syntax Command message FUNC BREAK RTIM time Query message FUNC BREAK RTIM Data time Data format float Data rang...

Page 71: ...rc upper current for breakdown test to 5mA Command message FUNC BREAK ARC 5 Query message FUNC BREAK ARC Return value 5 0 FUNC BREAKdown CONTinue To set inquiry about the CONTINUE ON OFF for voltage b...

Page 72: ...quiry about the TRIGGER DELAY for test Syntax Command message SYSTem MEA TRGDLY time Query message SYSTem MEA TRGDLY Data time Data format float Data range 0 0 99 9s Data accuracy 0 1s Data unit s Exa...

Page 73: ...message SYSTem MEA RPTINT time Query message SYSTem MEA RPTINT Data time Data format float Data range 0 0 99 9s Data accuracy 0 1s Data unit s Example Set the test interval to 1 0s Command message SYS...

Page 74: ...A STEPHOLD time Query message SYSTem MEA STEPHOLD Data time Data format float Data range 0 1 99 9s KEY Data accuracy 0 1s Data unit s Example Set the step hold interval to 1 0s Command message SYSTem...

Page 75: ...cters Data range OFF 0 ON 1 Example Set the auto rang to ON Command message SYSTem MEA AUTORANGE 1 Query message SYSTem MEA AUTORANGE Return value 1 SYSTem MEA GFI To set inquiry about the state for G...

Page 76: ...NV BEEPVOL 3 Query message SYSTem ENV BEEPVOL Return value 3 SYSTem ENV PASSVOL To set inquiry about the state for PASS VOLUM Syntax Command message SYSTem ENV PASSVOL ON OFF or 1 0 Query message SYST...

Page 77: ...ate Query message SYSTem ENV KEYLOCK Data state Data format characters Data range 0 1 0 manual 1 bus Example Set key lock state to manual Command message SYSTem ENV KEYLOCK 0 Query message SYSTem ENV...

Page 78: ...sage SYSTem ENV TIME hour minute second Query message SYSTem ENV TIME Data hour minute second Dataformat characters Data range 0 0 0 23 59 59 Example Set the system time to 16 23 23 Command message SY...

Page 79: ...al storage Syntax Command message MMEM COPY file name Data file name Data format characters Example Set copy file name to TH9110TEST Command message MMEM COPY TH9110TEST Return value copy pass return...

Page 80: ...copy fail return ERROR 5 2 7 FETCH Subsystem Commands The FETCh subsystem command is used to obtain the measurement results of the tester and is issued after the test is started until the end of the...

Page 81: ...Note The tester will automatically return each test result by default per step test result 5 2 8 Other Commands IDN Inquiry about the model and version information of the instrument Return message ma...

Page 82: ...0mA voltage 1 5kV Resolution 0 1uA Accuracy 1 reading 0 5 full scale Maximum Short Circuit Current 200mA AC test only Fast Discharge Function Auto discharge after test ends DCW Insulation Resistance T...

Page 83: ...hen both ends are short circuited high voltage delivery is allowed refer to Figure3 2 INTER LOCK Panel Operation Protection Key Lock Password Alarm Indication Pass short beep green light Fail long bee...

Page 84: ...onal and technical personnel to carry out please do not replace the internal components of the tester without authorization after the tester is repaired it needs to be re calibrated to avoid affecting...

Reviews: