![Tonghui Electronics TH9110 Operation Manual Download Page 70](http://html1.mh-extra.com/html/tonghui-electronics/th9110/th9110_operation-manual_1126462070.webp)
◇
70
Query message: FUNC:BREAK:LOWC?
Return value: 1.000
FUNC: BREAKdown:TTIM
To set /inquiry about the TEST TIME for voltage breakdown
test.
--
Syntax:
Command message: FUNC:BREAK:TTIM<time>
Query message: FUNC:BREAK:TTIM?
--
Data<time>:
Data format: float
Data range: 0, 0.3
~
999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the test time for breakdown test to1s.
Command message: FUNC:BREAK:TTIM 1
Query message: FUNC:BREAK:TTIM?
Return value: 1.0
FUNC: BREAKdown:RTIM
To set /inquiry about the RISE TIME for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:RTIM<time>
Query message: FUNC:BREAK:RTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the rise time for breakdown test to1s.
Command message: FUNC:BREAK:RTIM 1
Query message: FUNC:BREAK:RTIM?
Return value: 1.0
FUNC: BREAKdown:WTIM
To set /inquiry about the WAIT TIME for voltage breakdown
test.
--
Syntax:
Command message:FUNC:BREAK:WTIM<time>
Query message: FUNC:BREAK:WTIM?
--
Data<time>:
Data format: float
Data range: 0
~
999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s