5.1.2.2 GPIB Interface Function
This tester provides most of the general functions of GPIB except the controller, see the following
table:
Code
Function
SH1
Support all data source contact functions
AH1
Support all digital panel meter contact functions
T5
Basic talk function; Talk only function; Talking cancelled when MLA; Do
not support serial roll call
L4
Basic listen function; Listen cancelled when MTA; No listen only function
RL1
Remote/Local function
DC1
Device clearance function
DT1
Device trigger function
C0
No controller function
E1
Open-collector drive
5.1.2.3 GPIB Address
The GPIB of this tester is addressed by single address, no secondary address, 0-30 can be used as
the GPIB address, the factory default address is 1, the address value can be automatically saved in
Summary of Contents for TH9110