Syntax:
Command message: FUNC:BREAK:STEOVOLT<step voltage >
Query message: FUNC: :BREAK: STEOVOLT?
--
Data<step voltage >:
Data format: float
Data range: 5~500
Data accuracy: 1
Data unit: V
--
Example:
Set the step voltage for breakdown test to 50V.
Command message: FUNC:BREAK: STEOVOLT 50
Query message: FUNC:BREAK: STEOVOLT?
Return value: 50
FUNC:BREAKdown:MODE
To set /inquiry about the MODE for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:MODE<test mode>
Query message: FUNC: :BREAK: MODE?
--
Data<mode>:
Data format: integer
Data range: 0 ~ 1 (0: AC; 1: DC)
--
Example:
Set the mode for breakdown test: 1 (DC)
Command message: FUNC:BREAK:MODE 1
Query message: FUNC:BREAK: MODE?
Return value: 1
FUNC:BREAKdown:VOLTStart
To set /inquiry about the VOLTAGE START for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:VOLTS<voltage >
Query message: FUNC: :BREAK:VOLTS?
--
Data<voltage>:
Data format: float
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the starting voltage for breakdown test to 1000V.
Command message: FUNC:BREAK:VOLTS 1000
Query message: FUNC:BREAK:VOLTS?
Return value: 1000
FUNC:BREAKdown:VOLTEnd
To set /inquiry about the VOLTAGE END for voltage