Operation Manual
55
5.5.1.3. Glare Sensitivity Test
To conduct this test the additional lamps have to be
necessarily installed to the panel.
The lamps are optional equipment and are provided
separately – on request.
Installation instructions are supplied with the lamps.
In this test two lamps are situated both on the right and the left side of the chart
panel. The lamps are switched on at the beginning of the test and switched off
automatically after two minutes later. The lamp are provided separately.
The test should be conducted at the
distance of
1,5 meters
.
The backlight will be switched off when you choose the glare sensitivity test with
the OK button. If the OK button will be pressed again, then backlight will be
switched on and the test is starting, the patient should be asked, in binocular vision,
to recognize the letters in the first or second line.
For the licenses of group 1, for passing the test is required reading at least two-
thirds of the letters in the first line, corresponding to 0,1 in a maximum of one
minute. For the licenses of group 2 is required the reading in the same range of a
minute of the letters in the second line, corresponding to 0,2.