Bias may be applied in the following two ways:
•
Percentage:
A percentage of the actual value is calculated.
•
Plus/minus:
A compensating value is added to or subtracted from the actual value.
1. Touch the converted velocity key (1) on
the measurement window (fig. 1) to
access the setting window (fig. 2).
2. Enter the converted velocity by touch-
ing the number keys (2). The accept-
able range is 1,200 - 2,000 m/s.
In the event of an erroneous entry,
touch the D
De
elle
ette
e key (3).
3. Touch the R
Re
ettu
urrn
n key (4) to set the
entered value.
4. Select either percent bias (%) or
plus/minus bias (
µ
m) at the bottom of
the window (5).
5. When setting percent bias, enter the
bias rate by touching the number keys
(2). When setting the plus/minus bias,
enter the compensating value by
touching the number keys (2). The
acceptable ranges are:
Bias rate (%):
60 - 130%
Compensating value: -600 - +450
µ
m
In the event of an erroneous entry,
touch the D
De
elle
ette
e key (3).
6. Touch the R
Re
ettu
urrn
n key (4) to set the
entered values.
7. Touch the converted velocity key (1) to
return to the measurement window
(fig. 1).
T
Th
he
e b
biia
ass vva
allu
ue
e ((%
% o
orr
µ
m
m)) cca
an
n b
be
e cch
ha
an
ng
ge
ed
d a
afftte
err tth
he
e m
me
ea
ass--
u
urre
em
me
en
ntt h
ha
ass b
be
ee
en
n p
pe
errffo
orrm
me
ed
d..
Summary of Contents for AL-2000
Page 9: ...AL 2000 Biometer Pachymeter...
Page 13: ...AL 2000 Biometer Pachymeter 4...
Page 89: ...AL 2000 Biometer Pachymeter 80 8 8 M Me em mo or ry y C Ca ar rd d...
Page 111: ...AL 2000 Biometer Pachymeter...
Page 118: ...AL 2000 Biometer Pachymeter 109...
Page 119: ......