Measurement range
• Axial length:
15.00 - 40.00 mm
• Anterior chamber depth:
2.00 - 7.00 mm
• Lens thickness:
2.00 - 6.00 mm
>
>
Measurement accuracy and resolution
• Accuracy:
+0.1 mm
• Resolution:
0.01 mm
>
>
Converted velocity - factory setting and setting range
• Mean velocity of axial length
Normal:
1,550 m/s
1,500 - 1,600 m/s
Dense cataract:
1,548 m/s
1,500 - 1,600 m/s
Aphakic:
1,532 m/s
1,430 - 1,630 m/s
• Lens velocity
Normal:
1,641 m/s
1,540 - 1,740 m/s
Dense cataract:
1,629 m/s
1,540 - 1,740 m/s
• Anterior chamber velocity
Normal:
1,532 m/s
1,430 - 1,630 m/s
Dense cataract:
1,532 m/s
1,430 - 1,630 m/s
Aphakic:
1,532 m/s
1,430 - 1,630 m/s
• Velocity of vitreous body and anterior chamber
Pseudophakic:
1,532 m/s
1,000 - 2,000 m/s
• IOL velocity
Pseudophakic 1:
2,718 m/s
800 - 3,000 m/s
Pseudophakic 2:
1,049 m/s
800 - 3,000 m/s
Pseudophakic 3:
2,200 m/s
800 - 3,000 m/s
• IOL thickness
Pseudophakic:
1.00 mm
0.10 - 4.00 mm
>
>
IOL formulae
• SRK II
• SRK/T
• Holladay
• Hoffer-Q
• Haigis standard
• Haigis optimized
>
>
Biometry probe
• Type:
solid state
• Fixation light:
built-in red LED
• Transducer frequency:
10 MHz ±10%
• Tip radius:
5.0 mm (concave)
• Dimensions/weight:
8 mm x 100 mm/30 g