12.1 Explanations and countermeasures for error messages
12-4
Error classification
Error message
Explanations and countermeasures
No Switch Box
This message is indicated when switching box
is not recognized in multi-channel mode.
Confirm the connection of switching box.
Refer to
“9.2 Switching box” for the way of
connection. If you do not use any switching
box, set single-channel mode. Refer to
“4.4
Channel setting
” for the way of switching
channel mode.
The CF card may have been damaged.
Remove necessary data from the CF card, and
then format the CF card. If the problem
continues, use a new CF card.
Initial-in operation is prohibited (set to Disable)
in the setting. Refer to
“5.10 Measurement
auxiliary setting
” for the way of setting initial-in
permission.