![Tohnichi TDT3-G Operating Instruction Download Page 28](http://html1.mh-extra.com/html/tohnichi/tdt3-g/tdt3-g_operating-instruction_1124791028.webp)
Required weights
(1) For TDT60CN3-G: 1 piece of WS-TCL2
required.
(2) For TDT600CN3-G: 1 piece of WS-TCL2,
1 piece of WP-TCL2 and 1 piece of WP-
TCL5 required.
American Unit (
ozf·in
)
TDT60CN3-G
TDT600CN3-G
Measuring
Loading
Measuring
Loading
Point
Weights
Remarks
Point
Weights
Remarks
[ozf·in]
[kg]
[ozf·in]
[kg]
3
0.1080
Scale Pan + 8.0g
30
0.2160
Scale Pan + 116.0g
20
0.7200
Scale Pan + 620.0g
200
1.440
Scale 440g
40
1.440
Scale 440g
400
2.880
Scale 1.880kg
60
2.160
Scale 1.160kg
600
4.320
Scale 3.320kg
80
2.880
Scale 1.880kg
800
5.760
Scale 4.760kg
Components of TDT3-G Calibration Kit:
Calibration Scale Pan 100g
1 pc.
Calibration Scale Holder 1 kg
1 pc.
Calibration wire
1 pc.
Calibration lever
1 pc.
Calibration roller
1 pc.
Mounting bolt
2 pcs.