5-6. Deletion of Data
(1) Delete one data
Select the data you want to delete by using
st
key. Push C key to delete the data.
(2) Delete a range of data
Select the last counter of the data range you want to delete by using
st
key. Push STAT
key, the “Stt” appears on the display. Then select the first counter of the data range you want
to delete by using
st
key and push STAT key again. While the display shows “n”, “HI”, “Lo”
or “Av”, push STAT key and C key at the same time to delete the selected range of data.
(3) Delete all data
While pushing MEMORY/RESET switch, turn on the power. “CLEAr” appears on the display
and all the data will be deleted.
Attention
* Before deleting data, check again if the data is really OK to delete.
* Before turning the power off, make sure to transfer all the necessary data to an external device.
5-7. Statistical Processing Function (Number of sample, Max, Minimum, Mean value)
(1) Using
st
key, select the last counter of the data range you want to calculate.
(2) Push STAT key, then “Stt” appears on the display. Using
st
key, select the first counter of
your selected data range.
Push STAT key, then it shows the number of sample “n”, maximum value “HI”, lowest value
“Lo”, and the mean value “Av” in order.
5-8. Auto Zero Adjustment
Auto zero function will be activated when pushing C key or turning power off/on. Auto zero
function works only when the displayed value is below the values shown in the below table.
Remove the torque screwdriver from the tester when activating Auto Zero Adjustment.
When the torque load is above the value shown on the table, the display shows “Err9”.
<<In case Err9 appears>>
• Push C key at no load condition.
· If Err9 disappears, it functions properly.
• If Err9 does not disappear, turn off the power once, turn it on and try again.
· If Err9 does not disappear, the torque sensor or the circuit board may have error.
TDT60CN3
3 cN·m
TDT600CN3
30 cN·m