5-7. Statistical Processing Function, Number of samples, Max-Min-Mean Value
#1 Use ▼/▲ keys and show the last result of the section for statistical processing.
#2 Press “
STAT
” key to show “
Stt
” on display. And use ▼/▲ keys to show the first result of the section.
Press “
STAT
” key and the sampling number “
n
”, maximum value “
HI
”, minimum value “
Lo
” and mean
value “
Av
” in the specified data range are displayed in order.
5-8. Auto Zero Adjustment
With no load on the tester, Auto Zero function is performed automatically by press “
C
” key or turn the
power OFF/ON switch.
“Err9” on display
When loading torque exceeds more than 5% of the tester's maximum torque range, “
Err9
” is displayed.
When “
Err9
” is displayed, press “
C
” key with no load on the tester. “
Err9
” message disappears.
If “
Err9
” message is still on the display, turn off and on again.
Torque sensor or the electric circuit board may have any problem if “
Err9
” shows again.
5-9. Over-Torque Alarm
When loading torque exceeds 110% of its maximum torque range, the display flashes as a warning.
5-10. Error Display
DOTE4-G has a self-diagnosis function. When a problem occurs, any of the error messages from Err1
to 9 is displayed.
For any other error messages, refer to "Section #10. Error Message"