19
D
IGIT
AL T
ORQUE WRENCH TES
TER
D
O
TE4-G
8-2. PC Output
Connect the DOTE4-G and an external device with a communication cable
(Catalog No.383 or 385). Set the communication output format to “PC” and
ensure that other communication settings are consistent with the external device.
(See “5 Various Settings” for the setting methods.)
* Do not use an RS232C communication cable and a USB communication cable at the same time.
* Driver is required for communications with the No. 385 cable.
Install the driver from the CD-R accompanying the No. 385 cable.
• Communication settings
Data format
Transmission system
Serial Baud rate
Data length
Stop bit
Parity
: RS232C-compliant
: Start-stop synchronization
: 2400/4800/9600/19200 bps
: 7 bits/8 bits
: 1 bit
: None/Even number/Odd number
(1) Progressive output
The data are output when the "
Memory/Reset
" button is pressed or the auto memory/reset function is
activated.
(2) Batch output
Display the last memory count value in the output range and press "
STAT
" key.
“
Stt
” is displayed. Using ▼or ▲ key, display the first memory count value in the output range and
press "
STAT
" key. Then, press ▲ key while “
n
” is displayed.
The measured data of a selected range are collectively output.
The 7-digit of alphanumetric can be changed by command input.
Serial number of DOTE4-G is registered at shipment.
R E ,
0 1 2
3 ,
5
0
.
0 0 0
, n m
, A , O K ,
1 2
3
4
5 6 A
CR LF
R E ,
0 1
, 5
0
.
0 0 0
,
CR LF
Counter
4-digit
Direction
CW : " " blank
CCW :" - "
Torque value
6-digit
Torque unit
kgf·cm: kgfcm
kgf·m:kgfm
cN·m:cnm
N·m:nm
lbf·in:lbfin
lbf·ft:lbfft
Preset symbol
"A" to "J"
7-digit of alphanumetric.
Default : Serial number of the body
Counter 2-digit
Torque value
6-digit
PC Output Format
• For memory mode 1000
Judgment result
-Within preset Hi/Lo value: "OK"
-Higher than preset Hi value: "HI"
-Lower than preset Lo value: "LO"
-Judgment function is off: " " Blank
• For memory mode 99 (Compatible with the DOTE3-G)