16
D
IGIT
AL T
ORQUE WRENCH TES
TER
D
O
TE4-G
(13) Click type (QL, SP, etc):
Turn the handle of DOTE4-G till it reaches the set torque and "click".
Direct-reading type (CEM, DB, F, etc):
Turn the handle of DOTE4-G slowly till it approaches the measuring point.
(14) Click type (QL,SP, etc):
After reaching the measuring point and "click" sound, release the loading.
If you select manual operation, press "
MEMORY/RESET
" button to save the data,
then it goes on to the next counter value.
If you select AUTO MEMORY/RESET, it saves the data automatically after releasing
the load down to below 2% of the maximum torque of each measurement,
and forward the counter to the next.
(15) Direct-reading type (CEM, DB, F, etc):
When it approaches the measuring point, press "
MEMORY/RESET
" button to save the data,
and forward the counter to the next.
(16) Repeat steps (13), (14) and (15) to complete calibration.