11
D
IGIT
AL T
ORQUE WRENCH TES
TER
D
O
TE4-G
#3 Data length setting
Use ▲/▼ keys and select “
7bit
” or “
8bit
”. Press “
STAT
” key to enter and got to next setting,
“Parity Setting”. Press “
MD
” key to skip the setting. Press “
C
” key to go back to the measurement mode.
Factory
Default
Setting
:
8bits
#4 Parity setting
Use ▲/▼ keys and select “
nonE
” for none, “
EVEn
” for even number or “
odd
” for odd number.
Press “
STAT
” key to enter and go back to the measurement mode. Press “
MD
” key to skip the setting.
Press “
C
” key to go back to the measurement mode.
Factory Default Setting : None
Memory mode setting
Set a measured data memory capacity.
“
1000
”: 1,000 data memory
“
99
”: 99 data memory, communication format compatible with previous DOTE3-G model
Factory Default Setting : 1000
Use ▲/▼ keys and select 1000 or 99. Press “
STAT
” key to go confirmation screen.
Press “
MD
” key to skip the setting. Press “
C
” key to go back to the measurement mode.
Press “
STAT
” key to confirm. Memory mode is changed and all stored data is cleared.
When memory mode is changed, existed memory is cleared. Press “
C
” key to go back.
Default setting
Use ▲/▼ keys to select “
dFLt-Y
”. Press “
STAT
” key to r
eset to the factory default settings.
Press “
MD
” key to skip the setting. Press “
C
” key to go back to the measurement mode.