TOELLNER TOE 9261 Instruction Manual Download Page 20

 

Section 4 

– Technical Data 

 

 

TOELLNER Electronic Instrumente GmbH 

20 

(Technical data, cont.) 

General

 

 

Mains voltage 

Wide-range input 

90…264 V AC, 47-63 Hz, I

max

 = 0.4 A,  

rear appliance plug with fuse T1AL according to IEC 127-2/III, 
DIN 41662 

Power consumption, mains side 

Max. 35 VA 

Protection 

Protection Class I according to DIN EN 61140 

 

Degree of protection IP20 according to IEC 60529 

 

Overvoltage category II according to IEC 60664-1 

 

Pollution degree 2 according to IEC 60664-1 

Operating temperature 

0 - 40°C 

Storage temperature 

−20 - 70°C 

Reference temperature 

23°C ±1°C 

Cooling 

2-stage thermostatically controlled fan 

Overall dimensions (W x H x D) 

224 mm x 103 mm x 348 mm 

Rack mount dimensions (W x H x D) 

224 mm x 88 mm x 325 mm 

19" system 

Compatible with ½ 19", 2 HU 

Weight 

Approx. 4 kg 

Housing 

Aluminum/steel 

 
 
 

4.2 

General data 

 

 

Ordering 
options

 

  0-50 A, 0-60 V ..... TOE 9261-50 
0-100 A, 0-60 V ..... TOE 9261-100 

 

 

Included 
accessories

 

1 mains cable 
2 terminal blocks for signal line switches 
1 Instruction Manual 
 

Optional 
accessories

 

0.50 m connection cable with 1 safety socket, red ........................................... TOE 9260/22 
0.50 m connection cable with 1 safety socket, blue .......................................... TOE 9260/23 
1.20 m connection cable with 1 safety socket, red ........................................... TOE 9260/24 
1.20 m connection cable with 1 safety socket, blue .......................................... TOE 9260/25 
19" adapter set 2 HU (asymmetrical) for 1x TOE 9261 .................................... TOE 9521 
19" adapter set 2 HU (parallel connector) for 2 Toellner Instruments ½ 19" .... TOE 9522 
 
 
 

4.3 

Specification of the high-current connectors 

 
The 

built-in high-current connectors of the TOE 9261 are the following components

TOE 9261-50......... Safety flush-mounting plug, type ID/S6AR-N-B4S, manufacturer Multi-

Contact 

TOE 9261-100 ...... Safety flush-mounting plug, type ID/S6AR-N-S, manufacturer Multi-

Contact 

 
Suitable counter piece for crimping assembly on cables is the following: 
TOE 9261-50 and TOE 9261-100 ....... Safety socket, type KBT6AR-N/16-S, manufacturer 

Multi-Contact. 
This counter piece can only be attached to a cable by 
crimping assembly using special tooling, soldering is 
impossible. 

 

Summary of Contents for TOE 9261

Page 1: ...n Manual Electronic Switch TOE 9261 9261E Manual Rev05 doc R TOELLNER ELECTRONIC INSTRUMENTE GMBH Gahlenfeldstra e 31 58313 Herdecke Germany 49 0 23 30 97 91 91 Fax 49 0 23 30 97 91 97 E Mail info toe...

Page 2: ...switching a ground line minus 10 2 4 6 Connection of the control signal 10 2 4 7 Control of switching paths and operation of the consumer 10 2 4 8 Display of the switching states 11 2 4 9 Use of the b...

Page 3: ...ual When operating electronic equipment certain parts in these devices are necessarily under dangerous voltage Non observance of the warnings may therefore result in severe bodily injury or damage to...

Page 4: ...a feeding source to the connected consumer The branch path in this case is arranged parallel to the consumer and can be optionally connected via a DIL switch It serves as needed to selectively dischar...

Page 5: ...h TTL CMOS 5 V and LVCMOS 3 3 V Permissible switching frequency DC up to 100 kHz Protected against overcurrent short circuit excessive temperature excessive switching frequency 1 2 Model overview Seri...

Page 6: ...result in a hazardous condition for the device No intentional interruption of the protective earth connection is permitted Observe local regulations for earthing Ensure that only fuses of the specifi...

Page 7: ...or rear CONTROL input 3 Front panel BNC input CONTROL for controlling the switching state both for the power switch and the signal line switches 2 4 7 Control of switching paths and operation of the c...

Page 8: ...e illuminated 12 Red OVL LED Overload Comes on when a shutdown has occurred due to one of the following fault conditions Max switching frequency of 100 kHz on the control input has been exceeded Overc...

Page 9: ...f the consumer for switching a ground line the input voltage should be connected to this socket and to the Minus socket otherwise to the left hand Plus socket and the Minus socket The applied voltage...

Page 10: ...also the above mentioned improper load current flow via the 0 1 resistor is automatically locked Warning In the described switch configuration the positive supply line from the voltage source to the c...

Page 11: ...PULL UP ON to ON the internally evaluated level changes to High resulting in switching path S1 becoming conductive and switching path S2 non conductive In the configuration for switching the positive...

Page 12: ...e rise and fall times are lengthened High buffer capacities with low series resistance values ESR can also result in triggering of the overcurrent fuse Warning Feedback from the buffered load side to...

Page 13: ...ged by excessive ripple current loads 2 5 Operation of the signal line switches 2 5 1 Connection To switch signal lines they are routed through the bidirectional switches S3 S6 They are connected usin...

Page 14: ...reset is accomplished by turning the unit off and on using the mains switch The cause of the overcurrent or short circuit must be identified and eliminated 2 6 2 Input overcurrent protection The input...

Page 15: ...n responds to excessive component temperatures in the unit This protection is provided in two stages For continuous operation at high voltage high current and high switching frequency the fan speed is...

Page 16: ...to the 19 adapter piece 1 Insert this unit in replacement for the fill strip and screw it on to the front panel using the hexagon socket screws 6 4 For the other side attach the adapter bracket 3 to...

Page 17: ...ice Pull the four fill strips 5 from the side 3 For the outer faces of the devices attach a handle 4 to a 19 adapter piece 1 each using the countersunk screws 7 Insert these units in replacement for t...

Page 18: ...ower switch switching times Rise time tr with reference resistor 1 With connected input buffer capacitor Without buffer capacitor input cable 0 50 m twisted Without buffer capacitor input cable 1 20 m...

Page 19: ...Increase in fan speed Stage 2 Switching off of switching paths display OVL Switching off and on again using mains switch following appropriate cooling period Power switch is not reverse polarity prot...

Page 20: ...luded accessories 1 mains cable 2 terminal blocks for signal line switches 1 Instruction Manual Optional accessories 0 50 m connection cable with 1 safety socket red TOE 9260 22 0 50 m connection cabl...

Page 21: ...t 8 Selectable base load 10 k or 100 k with green indication LED 9 Power switch branch path incl red indication LED 10 0 series resistor for branch path with green indication LED 11 0 1 series resisto...

Reviews: