2.2  

Connection to Load Cell

 

 

Description 

No. 

Description 

Code 

Positive excitation 

EXC+ 

Negative excitation 

EXC- 

Positive signature 

SIG+ 

Negative signature 

SIG- 

5

 

Positive feedback 

SN+ 

Negative feedback 

SN- 

Shield cable 

SHIELD 

If use four wire shielded cable, please make short circuit of plus excitation and plus feedback, make short 

circuit of minus excitation and minus feedback!!! 

Load cell connect with indicator must be reliable; shield wire must be connected to ground reliably. 

Load cell and indicator are static sensitive devices, taking measures to prevent from static power when using.

 

During the thunder storm season, proper lightning protection should be taken to protect the load cell and 

indicator from damaging by lighting and to ensure the personal safety to run of the weighing and related 

equipment. 

 
2.3 Connection to printer 

 

Instruction

 

No 

Description 

Code 

data strobe signal 

STB 

8 digits parallel data   

D0 

D1 

D2 

D3 

D4 

D5 

D6 

D7 

11 

“BUSY”signal 

BUSY 

25 

GROUND 

GND 

 

Print note: 

Print function can be used after set-up. 

Indicator print port output cable must be connected to printer correctly, special printer cable is necessary.

 

Due  to various printer, and different characteristics, please select recommended printer.

    

 

Printer must connect to ground reliably,  otherwise it  may influence the indicator normal use,or  broke the load 

cell ,indicator,scoreboard and printer. 

 
 
 
 
 
 
2.4 Connection to soreboard 

Summary of Contents for TTA-01

Page 1: ...TIGER TTA 01 Weighing Indicator User manual Version Oct 2015...

Page 2: ...gas or flammable vapor areas or tank system with pressure keep indicator and load cell away from strong electric magnetic field corrosive substances and explosive materials do not sue strong solvents...

Page 3: ...istance 30 meters Serial communication interface Transmission communication interface RS232 Transmission distance 30 meters Baud rate 1200 2400 4800 9600 19200 is optional Printer interface standard p...

Page 4: ...hunder storm season proper lightning protection should be taken to protect the load cell and indicator from damaging by lighting and to ensure the personal safety to run of the weighing and related eq...

Page 5: ...scription Code 5 Connect to ground GND 1 RS485 A A 9 RS485 B B 3 RS232 receive RXD 2 RS232 send TXD Connect communication interface and PC side correctly or will break the indicator interface and PC s...

Page 6: ...eters of communication address baud rate communication mode 4 2 2 Print Settings Go into this menu it could set print type print format vehicle number and cargo number 4 2 3 Time Settings If the date...

Page 7: ...r Print the Report function if find there is a symbol like it means the printing data is not complete Appendix 1 serial communication data format All data is ASCII code and each byte could be 8bits da...

Page 8: ...ght the format is below Each frame is made up of 14 groups data a symbol or b h net weight value 6ofnumber and 1 decimal NOTICE XOR verify high four bits and low four bits confirme like below XOR high...

Page 9: ...C N GROSS TARE NET 0001 12 01 08 20 08 00 ZJ888FUK APPLE PIE 8000 1300 6700 0002 12 01 08 20 19 21 32D AK047 ORANGE 9000 1300 7700 0003 12 01 08 20 24 33 BMW86 SB BANANA 8600 1500 7100 Appendix 3 Fun...

Page 10: ......

Page 11: ...ing list Serial no Name and specifications Quan tity Remark 1 indicator 1pc 2 User manual 1pc 3 Technical manual 1pc 4 Certificate 1pc 5 7 core aviation plug 1pc 6 insurance tube 0 5A 1pc 7 Double AC...

Reviews: