![TiePie Handyscope HS5 series User Manual Download Page 39](http://html1.mh-extra.com/html/tiepie/handyscope-hs5-series/handyscope-hs5-series_user-manual_1114538039.webp)
9.3
Trigger system
System
Digital, 2 levels
Source
CH1, CH2, digital external, OR,
generator start, generator new period, generator stop
Trigger modes
Rising edge, falling edge, any edge,
inside window, outside window,
enter window, exit window,
pulse width
Level adjustment
0 to 100% of full scale
Hysteresis adjustment
0 to 100% of full scale
Resolution
0.024 % (12 bits)/0.006 % (14/16 bits)
Pre trigger
0 to 64 MSamples measuring 1 channel,
0 to 32 MSamples measuring 2 channels,
1 sample resolution
Post trigger
0 to 64 MSamples measuring 1 channel,
0 to 32 MSamples measuring 2 channels,
1 sample resolution
Trigger hold-off
0 to 64 MSamples, 1 sample resolution
Trigger delay
0 to 16 GSamples, 1 sample resolution
Segmented trigger
Available via
Maximum number of segments
1024
Minimum segment length
1 sample
Maximum segment length
64 M / number of segments measuring 1 channel
32 M / number of segments measuring 2 channels
Trigger rearm time
Sample frequency dependent,
<
700 ns on highest sample frequency
Digital external trigger
Input
Extension connector pins 1, 2 and 3
Range
0 to 2.5 V (TTL)
Coupling
DC
Jitter
Depending on source and sample frequency
Source = channel
≤
1 sample
Source = external or generator
Sample frequency = 500 MS/s
≤
8 samples
Sample frequency
<
500 MS/s
≤
4 samples
Sample frequency
≤
100 MS/s
≤
1 sample
Specifications
35
Summary of Contents for Handyscope HS5 series
Page 1: ...Handyscope HS5 User manual TiePie engineering...
Page 6: ...2 Chapter 1...
Page 17: ...Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated Introduction 13...
Page 18: ...14 Chapter 3...
Page 24: ...20 Chapter 4...
Page 28: ...24 Chapter 5...
Page 30: ...26 Chapter 6...
Page 32: ...28 Chapter 7...
Page 46: ...42 Chapter 9...