When measuring DC signals, make sure to set the signal
coupling of the input to DC.
3.5
Probe compensation
The Handyscope HS5 is shipped with a probe for each input chan-
nel. These are 1x/10x selectable passive probes. This means that
the input signal is passed through directly or 10 times attenuated.
When using an oscilloscope probe in 1:1 the setting, the
bandwidth of the probe is only 6 MHz. The full bandwidth
of the probe is only obtained in the 1:10 setting
The x10 attenuation is achieved by means of an attenuation
network. This attenuation network has to be adjusted to the oscil-
loscope input circuitry, to guarantee frequency independency. This
is called the low frequency compensation. Each time a probe is
used on an other channel or an other oscilloscope, the probe must
be adjusted.
Therefore the probe is equiped with a setscrew, with which the
parallel capacity of the attenuation network can be altered. To
adjust the probe, switch the probe to the x10 and attach the probe
to a 1 kHz square wave signal. Then adjust the probe for a square
front corner on the square wave displayed. See also the following
illustrations.
12
Chapter 3
Summary of Contents for Handyscope HS5 series
Page 1: ...Handyscope HS5 User manual TiePie engineering...
Page 6: ...2 Chapter 1...
Page 17: ...Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated Introduction 13...
Page 18: ...14 Chapter 3...
Page 24: ...20 Chapter 4...
Page 28: ...24 Chapter 5...
Page 30: ...26 Chapter 6...
Page 32: ...28 Chapter 7...
Page 46: ...42 Chapter 9...