![TiePie Handyscope HS5 series User Manual Download Page 36](http://html1.mh-extra.com/html/tiepie/handyscope-hs5-series/handyscope-hs5-series_user-manual_1114538036.webp)
Pin 8, Power OUT, has the same potential as the Handyscope
HS5 power supply. When the Handyscope HS5 is USB powered,
it is at USB power level. When the Handyscope HS5 is externally
powered, it is at the same level as the external power input.
8.4
AUX I/O
The Handyscope HS5 has two Auxiliary I/O connectors at the rear
of the instrument. These can be used to combine multiple instru-
ments to a single combined instrument to perform synchronized
measurements.
Figure 8.6: Auxiliary I/O connector
Pin
Description
Pin
Description
1
GND
11
Data OK P EXT (LVDS)
2
EXT CLK P (LVDS)
12
Data OK N EXT (LVDS)
3
EXT CLK N (LVDS)
13
reserved
4
GND
14
Generator Trigger (I/O)
5
Data OK (I/O)
15
reserved
6
reserved
16
reserved
7
GND
17
GND
8
Ext Trigger (I/O)
18
reserved
9
reserved
19
GND
10
GND
Table 8.3: Pin description Auxiliary I/O connector
The I/O signals (pins 5, 8 and 14) can be used as input and
output. They are digital signals switching between 0 V and 2.5 V.
The LVDS external clock (pins 2 and 3) must be 10 MHz,
±
1%.
The Auxiliary I/O connectors use HDMI type C sockets, but
are not HDMI compliant. They can not be used to connect
an HDMI device to the Handyscope HS5.
32
Chapter 8
Summary of Contents for Handyscope HS5 series
Page 1: ...Handyscope HS5 User manual TiePie engineering...
Page 6: ...2 Chapter 1...
Page 17: ...Figure 3 6 correct Figure 3 7 under compensated Figure 3 8 over compensated Introduction 13...
Page 18: ...14 Chapter 3...
Page 24: ...20 Chapter 4...
Page 28: ...24 Chapter 5...
Page 30: ...26 Chapter 6...
Page 32: ...28 Chapter 7...
Page 46: ...42 Chapter 9...