3.2 commands
Temperature module TED8xxx / page 46
3.2 commands
3.2.1 Select the module slot
":SLOT <NR1>"
Selects a slot for further programming
<Nr1>=1…8 (PRO8000), 1…2 (PRO800)
":SLOT?"
Queries the selected slot
[:SLOT <NR1><LF>]
3.2.2 Thermistor calibration (exponential method)
(Not for TED8xxx with Kryo-option!)
Programming:
":CALTB:SET <NR3>"
Program the energy constant B
val
":CALTR:SET <NR3>"
Program the nominal resistance R
0
":CALTT:SET <NR3>"
Program the nominal temperature T
0
Reading:
":CALTB:SET?"
Reading the energy constant B
val
[:CALTB:SET
<NR3><LF>]
":CALTR:SET?"
Reading the nominal resistance R
0
[:CALTR:SET
<NR3><LF>]
":CALTT:SET?"
Reading the nominal temperature T
0
[:CALTT:SET
<NR3><LF>]
":CALTB:MIN?"
Reading the minimum B
val
allowed
[:CALTB:MIN
<NR3><LF>]
":CALTR:MIN?"
Reading the minimum R
0
allowed
[:CALTR:MIN
<NR3><LF>]
":CALTT:MIN?"
Reading the minimum T
0
allowed
[:CALTT:MIN
<NR3><LF>]