![THORLABS TED8020 Operation Manual Download Page 47](http://html.mh-extra.com/html/thorlabs/ted8020/ted8020_operation-manual_1110715047.webp)
2.3 Switching on and off
Temperature module TED8xxx / page 40
2.3 Switching on and off
First select the TED module in the main menu (the LED “SEL” must light)
#
(Refer to chapter 2.1.2, "Selecting a module" on page 27)
!
Attention
!
Before switching on a temperature module TED8xxx first set the TEC
limit current I
LIM
(hardware limit) for the applied TEC element with a
screwdriver.
The corresponding potentiometer is marked I
LIM
and is situated at the
front panel of the current module.
#
(Refer to chapter 1.6.1, "Setting the limit values" starting on page 13)
Pressing the key
will switch the module on or off not regarding the menu you
are in as long as the module is selected. The LED “ON” at the respective module will
light up when switched on.