39
5.3.2
Wiring and Software Settings for Hub Operation
Using the KCH601 hub greatly simplifies system wiring because the hub provides all
the necessary power, signal and USB comms connections to all controller cubes.
The following procedure describes a typical set up, with the KPA101 position aligner
cube being used with a pair of KPZ101 piezo driver cubes on the KCH601 controller
hub - see Fig. 5.2 for a general system schematic diagram. The K-Cube positions on
the hub and associated parameter settings described are by n o means the only
possible options. For further information, see Section 5.3.3.
1) Perform the mechanical installation as detailed in Section 3.2. of this manual and
also in the manual supplied with the K-Cube Piezo drivers.
2) Install the K-Cube units onto th e controller hub. Fit a piezo unit in bay 1, the
position aligner in bay 2 and the other piezo unit in bay 3, as shown in Fig. 5.2.
3) Connect the HV OUT terminals on the rear of the Piezo drivers to the associated
piezo actuators, Bay 1 to Y-axis and Bay 3 to X-axis.
4) Connect the laser to the laser driver.
5) Install any mechanical components such as turning mirrors etc. wh ich may be
required in the beam path. - see Fig. 5.2 for a typical set up..
Fig. 5.2 Typical Set Up
PC
HeNe
ASM003
Piezo Mount
Mirror Mount
PDQ80A QDP Sensor
Beam
Splitter
To Experiment
KPZ101
(bay 3)
KPZ101
(bay 1)
KPA101
(bay 2)
Position Aligner
MENU
Piezo Controller
MENU
Piezo Controller
MENU