© 2018 Thorlabs
20
CLD101x
2.5.1 Set Point Memories
Up to 4 set points (optical power or laser current and temperature) can be stored. Simply touch and
hold for about 3 sec a set point memory icon. A beep, accompanied by the message "Set points
stored" confirms saving.
Note
In case the memory is not empty, the data will be overwritten!
A set point can be retrieved by short touch of the icon, settings immediately take place. If the
memory data are related to a different operating mode, the laser must be switched off before the
CLD101x can load the set point. If the laser is on, an appropriate warning message appears.
Set point memories can be cleared only at once from the
Clear Memories
button in
panel.
2.5.2 Main Panel Status Messages
LD current output is locked
on the rear panel to
"enable" position
Interlock circuit is open
Close interlock circuit, e.g. by installing the
jumper
active
Laser switched off or paused due to exceeding
the given temperature window. Make sure the
TEC controller is ON.
reached
The laser current set value reached the limit.
reached
May appear for a short time after switching the
TEC on.
Actual temperature exceeded
.
Switch on TEC.
Device is in Remote Mode
An error in remote mode occurred. There are errors in the instrument's error queue.
Use the 'SYSTem:ERRor'
to read out the error queue
.
29
7
7
22
22
25
22
48
Summary of Contents for CLD101 Series
Page 1: ...Compact Laser Diode Controller CLD101x Operation Manual 2018 ...
Page 2: ...Version Date 1 7 09 Jul 2018 Copyright 2018 Thorlabs ...
Page 76: ... 2018 Thorlabs 74 CLD101x 6 2 Dimensions CLD101x ...
Page 84: ... 2018 Thorlabs 82 CLD101x 6 5 Certifications and Compliances ...
Page 90: ......
Page 91: ......
Page 92: ...www thorlabs com ...