2021-04
Additional pad elements (Fig. 46)
The number of pads fitted to the tGo can be adjusted according to individual requirements. To fit additional pads, the
black plastic element – that has a slot in it for the belt – must first be removed. To do this, remove the screw that con-
nects the outer pad with the plastic element (the supplied Allen wrench and a 10 mm ring or open end wrench are
required).
Pull the black plastic element out of the pad. Push the desired number of pads onto the pad ring. Ensure that the black
plastic pad holders are always pointing away from the child’s body. Always secure the pads using the supplied nuts and
screws. Finally, replace the belt guide components onto the pad ring and secure in-place.
Follow the same procedure for removing pads that are not required.
Tube for breast pads
The height of the breast pads can be individually adjusted on the tube.
Technical Data
t
Go
Size 1
Size 2
Seat height
25 cm - 60 cm / 9.8 - 23.6"
25 - 80 cm / 9.8 - 31.5"
Overall dimensions
83 x 54 cm / 32.7 x 21.3"
92 x 65 cm / 36.2 x 25.6"
Saddle length
20 cm / 7.9"
23,5 cm / 9.3"
Handgrip height
43 - 90 cm / 16.9 - 35.4"
45 - 110 cm / 17.7 - 43.3"
Pelvis support height
40 - 75 cm / 15.8 - 29.5"
40 - 95 cm / 15.8 - 37.4"
Wheel size - front
15 cm / 5.9"
15 cm / 5.9"
Wheel size - rear
20 cm / 7.9"
20 cm / 7.9"
Weight
12.4 kg / 27.3 lbs
13.8 kg / 30.4 lbs
Max. user weight
40 kg / 88.2 lbs
45 kg / 99.2 lbs
Material Specifications
Aluminum, stainless steel, steel (painted), plastic, rubber, polyester, polyurethane
Cleaning, Maintenance, and Repair
Cleaning, Care Instructions, and Disinfection on a Daily Basis
Hook and Loop Fasteners
To keep the hook and loop fasteners working well, clean them now and then with a brush. Always close or cover the
hook and loop fasteners before washing.
Plastic Parts
The plastic parts on the product can be cleaned and disinfected with standard soap (neutral soap).
Metal Parts
The metal parts on the product can be cleaned and disinfected with standard soap (neutral soap).
Disinfection
Clean the padding, plastic parts, and metal parts before disinfecting.
Damp wipe all the plastic and metal parts on the product with disinfectant. Only water-based products should be used.
When disinfecting note the directions from the manufacturer in question.
Instructions for Use –