![ThermoFisher Scientific Wellwash 4 Service Manual Download Page 78](http://html1.mh-extra.com/html/thermofisher-scientific/wellwash-4/wellwash-4_service-manual_1105610078.webp)
Chapter 8: Troubleshooting
Thermo Scientific Wellwash
®
4 Mk 2 User Manual Rev.2.3, Cat. no. 1507340
54
8.4
Defects and abnormal stresses
Whenever it is likely that the protection has been impaired, the instrument
should be made inoperative and be secured against any unintended operation.
The protection is likely to be impaired if, for example, the instrument:
1.
Shows any visible damage
2.
Fails to perform the intended functions
3.
Has been subjected to prolonged storage under unfavorable conditions
4.
Has been subjected to severe transport stresses
8.5
Before operation
Read this manual in its entirety, as it contains information necessary to ensure
safe operation.
8.5.1
Installation
The instrument should be unpacked and installed in accordance with Chapter 5.
Particular attention should be paid to the correct fitting of the tubing to obviate
leaks (Fig. 5.11). Always ensure that the electrical supply in the laboratory
conforms to that specified on the voltage label on the rear panel of the pump
unit.
8.5.2
In operation
Ensure that the waste bottle is emptied regularly and at least every time that a
wash bottle is loaded. The liquid trap bottle that prevents liquid from entering
the pump should be kept dry at all times.
8.6
Cleanliness
Clean the instrument regularly as stated below.
1.
It is recommended that the case of the instrument is cleaned periodically to
maintain its good appearance. It is particularly essential that the carrier
slideway be clean and dry to prevent jamming. A soft cloth dampened in
warm, mild detergent solution will suffice.
2.
If any surfaces have been contaminated with biohazardous material, a mild
sterilizing solution should be used. See Section 7.3 Decontamination
procedure.
Caution:
DO NOT use any solutions containing hypochlorite, such as bleach, on
any of the stainless steel surfaces, as this may cause permanent damage
to the finish.
Summary of Contents for Wellwash 4
Page 2: ......
Page 3: ......
Page 4: ...to ion...
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ...ressure on Vessel 7...
Page 9: ...imensisning o f t e elcctm icdlnsulatio nners 1e Allen K...
Page 10: ...nor its agents can b e carried out by any suita...
Page 11: ......
Page 12: ...arrier onto the stainless steel sli the last row to ive...
Page 13: ...initiates the was icates a vacuuln uttons is confirmed...
Page 15: ...e last wasl ing cycle oft1...
Page 16: ......
Page 17: ...ocedure i s requir Disinfection is not 1 g of the instrument...
Page 18: ......
Page 20: ...rotedon has been im e ainst any unintende...
Page 21: ...roblems refer to as often as necessa...
Page 22: ...nit witc Air leaks...
Page 23: ...ubing leaks e attention or if re contact...
Page 24: ......
Page 25: ...Wellwash Rev 2 3 Cat no 1507340 Thermo Scientific Wellwash 4 Mk 2 User Manual...
Page 92: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 68...
Page 93: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 69...
Page 94: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 70...
Page 95: ......
Page 96: ......
Page 97: ...FACE...
Page 98: ...Pines...
Page 99: ......
Page 100: ......
Page 101: ...e sensors e...
Page 102: ......
Page 103: ......
Page 104: ...Unit Filter Ins...
Page 105: ......
Page 106: ...ACT P...
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ...Only for old WW4 ONLY FOR OLD WW4 MODELS...
Page 111: ...ONLY FOR OLD WW4 MODELS...
Page 112: ...ONLY FOR OLD WW4 MODELS...
Page 113: ...1 i t c 1 itc t i n ONLY FOR OLD WW4 MODELS...
Page 118: ......
Page 119: ......
Page 132: ......
Page 134: ......
Page 139: ......
Page 144: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......