Chapter 6: Operation
Thermo Scientific Wellwash
®
4 Mk 2 User Manual Rev.2.3, Cat. no. 1507340
35
6.6
Resetting
If you press the RESET button on the keypad at any stage, this will stop the
immediate washing sequence and reset the system to its pre-programmed
state.
6.7
Shutdown procedure
Warning:
Remove any microplate still on the plate carrier. Dispose of all
microplates and strips as biohazardous waste.
The wash reagents may contain buffer salts, which can crystallize when the
washer is not in use. This may cause clogging of the wash/aspirate tips. To
avoid this, the following procedures should be followed.
6.7.1
Daily shutdown procedure
When the last wash cycle of the day is complete, switch the power OFF through
the switch at the rear of the pump unit. Fill the priming reservoir of the plate
carrier with distilled water and lower the wash head tips into the priming
reservoir. Do not detach the tubes unless you first open the wash bottle cap
slightly to release the pressure.
Wipe the instrument surfaces with a soft cloth or tissue paper moistened with
deionized distilled aqua, a mild detergent (SDS, sodium dodecyl sulfate) or soap
solution.
If you have spilt infectious agents on the instrument, disinfect with 70% alcohol
or other disinfectant. See the Wellwash 4 Mk 2 User Manual, Section 7.3.
Last of all, put the dust cover on.
6.7.2
Long-period shutdown
If the washer is not likely to be used for a long period, you should flush the
system through with distilled water prior to carrying out the DAILY shutdown
procedure. Empty and clean the waste bottle and check that the liquid trap is
dry.
When the washer is to be used after a prolonged shutdown, you should prime
the system (Section 6.3) at least four (4) times prior to carrying out any wash
function.
6.8
Emergency situations
In case there is any abnormal situation during operation, such as fluids spilling
inside the instrument, follow these steps:
1.
Switch OFF the instrument.
2.
Unplug the instrument from the mains.
3.
Perform appropriate corrective actions. However, do not touch the interior
of the instrument.
4.
Contact authorized technical service or your local Thermo Fisher Scientific
representative.
Summary of Contents for Wellwash 4
Page 2: ......
Page 3: ......
Page 4: ...to ion...
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ...ressure on Vessel 7...
Page 9: ...imensisning o f t e elcctm icdlnsulatio nners 1e Allen K...
Page 10: ...nor its agents can b e carried out by any suita...
Page 11: ......
Page 12: ...arrier onto the stainless steel sli the last row to ive...
Page 13: ...initiates the was icates a vacuuln uttons is confirmed...
Page 15: ...e last wasl ing cycle oft1...
Page 16: ......
Page 17: ...ocedure i s requir Disinfection is not 1 g of the instrument...
Page 18: ......
Page 20: ...rotedon has been im e ainst any unintende...
Page 21: ...roblems refer to as often as necessa...
Page 22: ...nit witc Air leaks...
Page 23: ...ubing leaks e attention or if re contact...
Page 24: ......
Page 25: ...Wellwash Rev 2 3 Cat no 1507340 Thermo Scientific Wellwash 4 Mk 2 User Manual...
Page 92: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 68...
Page 93: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 69...
Page 94: ...Thermo Scientific Wellwash 4 Mk 2 User Manual Rev 2 3 Cat no 1507340 70...
Page 95: ......
Page 96: ......
Page 97: ...FACE...
Page 98: ...Pines...
Page 99: ......
Page 100: ......
Page 101: ...e sensors e...
Page 102: ......
Page 103: ......
Page 104: ...Unit Filter Ins...
Page 105: ......
Page 106: ...ACT P...
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ...Only for old WW4 ONLY FOR OLD WW4 MODELS...
Page 111: ...ONLY FOR OLD WW4 MODELS...
Page 112: ...ONLY FOR OLD WW4 MODELS...
Page 113: ...1 i t c 1 itc t i n ONLY FOR OLD WW4 MODELS...
Page 118: ......
Page 119: ......
Page 132: ......
Page 134: ......
Page 139: ......
Page 144: ......
Page 149: ......
Page 150: ......
Page 151: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......